Use of the Hildebrand grid nebuliser for analysis of high matrix solutions containing easily ionisable elements with inductively coupled plasma mass spectrometry

Authors:
  • TBTimothy J. Brotherton
    Timothy J. Brotherton
  • WSWei L. Shen
    Wei L. Shen
  • JCJoseph A. Caruso
    Joseph A. Caruso

Abstract
Aqueous, high dissolved salt solutions and biological fluids (urine) were introduced with the Hildebrand grid nebuliser for inductively coupled plasma mass spectrometry (ICP-MS) detection. Optimisation studies of ICP power, plasma sampling position and nebuliser gas flow-rate were performed. Three-dimensional plots that show optimum system performance were obtained at a plasma sampling position of 12 mm, ICP power of range 1.35–1.50 kW and a nebuliser gas flow of 0.65 l min–1. These results are similar to previously reported data. Ionisation suppression as a function of synthetic ocean water (SOW) and urine concentration was studied. A 10% analyte suppression (Co, In, Ba, Pb) generally occurred for a 0.01–0.10 M SOW solution. The ionisation suppression for a 100% urine matrix is approximately 40%(Co, As, Cd, Ba, Pb). Detection limits for SOW were degraded by up to one and a half orders of magnitude. For the urine matrix, detection limits were degraded 15–40% for the elements studied.