Efficacy of an aerosol suction device Free-100 M in removing aerosols produced by coughing to minimize COVID-19 infection
Authors:
Open Access
- 30 November 2022
- journal article
- research article
- Published by Springer Nature in Journal of Anesthesia
- Vol. 37 (2) , 196-200
Abstract
Purpose The healthcare workers are at the greatest risk of being exposed to viral infection during airway management of a patient with coronavirus disease 2019 (COVID-19). An air extractor which suctions air around the patient's face would reduce the spread of viral aerosols during coughing, but no study has confirmed this. We assessed whether or not an air extractor reduces the amount of aerosols spreading toward the operator's face, during coughing of simulated patients. Methods After obtained approval of the study by a research ethics committee and written informed consent from 20 volunteers (and additional 20 volunteers), we asked each volunteer to lie supine on a table in a positive-pressure management operating room. As a cross-over design, we used an airborne particle counter (Handheld 3016, SGY company, Tokyo) to measure the aerosols approximately 30 cm above the participant's mouth, while the volunteer was coughing, with and without the use of an air extractor Free-100 M (Forest-one, Funabashi), facing the participant's mouth. In another 20 volunteers, the aerosols were measured, while each volunteer was lying supine, without coughing, and without the use of the air extractor. Results The aerosol count during coughing was significantly lower when the air extractor was used [median: 55 (interquartile range: 15–128)] than when it was not used [73 (44–201)] [p = 0.001, difference: 19 (95%CI: 4–70)]. Conclusions The Free-100 M air extractor would reduce, but do not remove all, aerosols produced by coughing of a patient, and thus may reduce the risk of infection of COVID-19.Keywords
This publication has 10 references indexed in Scilit:
- Aerosol extractor for airway management of COVID-19 patientsJournal of Anesthesia, 2021
- COVID-19: accurate interpretation of diagnostic tests—a statistical point of viewJournal of Anesthesia, 2020
- Aerosol containment device for airway management of patients with COVID-19: a narrative reviewJournal of Anesthesia, 2020
- Use of a high-flow extractor to reduce aerosol exposure in tracheal intubationBritish Journal of Anaesthesia, 2020
- Videolaryngoscopy for tracheal intubation in patients with COVID-19British Journal of Anaesthesia, 2020
- Measurement of airborne particle exposure during simulated tracheal intubation using various proposed aerosol containment devices during the COVID‐19 pandemicAnaesthesia, 2020
- Airtight, flexible, disposable barrier for extubationJournal of Anesthesia, 2020
- Anesthesia in the times of COVID-19Journal of Anesthesia, 2020
- Relationship between expired lung volume, peak flow rate and peak velocity time during a voluntary cough manoeuvreBritish Journal of Anaesthesia, 1994
- The Test That a Proportion Is .50 and the Sign Test1977