Performance evaluation of a turbidimetric cystatin C assay on different high-throughput platforms
Authors:
- Anders GrubbAGAnders Grubb
- Anders LidénALAnders Lidén
- Mats FlodinMFMats Flodin
- Annacarin BerggrenABAnnacarin Berggren
- Joris DelangheJDJoris Delanghe
- Veronique StoveVSVeronique Stove
- Hilmar LutheHLHilmar Luthe
- Karl-Heinz RhodeKRKarl-Heinz Rhode
- Claus BeckCBClaus Beck
- Ingrid DomkeIDIngrid Domke
- 17 May 2010
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 70 (5) , 347-353
Abstract
The goal with this study was to evaluate the analytical performance of a new cystatin C immunoassay (Tina-quant a Cystatin C, Roche Diagnostics GmbH). The evaluation was carried out at four centers according to a standardized protocol. The Tina-quant a Cystatin C is a latex particle-enhanced immunoturbidimetric assay. Roche cobas 6000, MODULAR ANALYTICS SWA and COBAS INTEGRA instruments were included in the study. Method comparison studies were carried out against two turbidimetric methods (Dako Cystatin C, Gentian Cystatin C), and one nephelometric method (Siemens N-Latex Cystatin C). Linearity was proven throughout the measuring range from 0.4 to 8 mg/L. Within-run CVs ranged from 0.7-2.8%, and total CVs from 1.4-4.7 % (concentration range 0.6-3.9 mg/L). Comparable results were obtained with paired serum and Li-heparinate plasma samples. Good agreement was achieved in the comparisons between the Tina-quant a Cystatin C assay and the other commercially available cystatin C assays, two different turbidimetric methods (slope range 0.88-1.04, intercept < 0.17 mg/L, r > or = 0.993) and one nephelometric assay (slope range 0.90-1.05, intercept < 0.21 mg/L, r > or = 0.986). The Tina-quant a Cystatin C assay was shown to be precise and accurate with proven linearity over the measuring range. Good comparability was obtained with other commercially available assays for the determination of cystatin C. The Tina-quant a Cystatin C assay is very well suited for clinical use on routine clinical chemistry analysers to detect renal dysfunction with a 24 h availability.Keywords
This publication has 30 references indexed in Scilit:
- Estimating GFR Using Serum Cystatin C Alone and in Combination With Serum Creatinine: A Pooled Analysis of 3,418 Individuals With CKDAmerican Journal of Kidney Diseases, 2008
- Expressing the Modification of Diet in Renal Disease Study Equation for Estimating Glomerular Filtration Rate with Standardized Serum Creatinine ValuesClinical Chemistry, 2007
- Cystatin C and Estimates of Renal Function: Searching for a Better Measure of Kidney Function in Diabetic PatientsClinical Chemistry, 2007
- Evaluation of Gentian cystatin C reagent on Abbott Ci8200 and calculation of glomerular filtration rate expressed in mL/min/1.73 m2from the cystatin C values in mg/LScandinavian Journal of Clinical and Laboratory Investigation, 2007
- Measurement of renal function in chronic renal diseaseKidney International, 1990
- Serum Creatinine and Renal FunctionAnnual Review of Medicine, 1988
- Serum Concentration of Cystatin C, Factor D and β2‐Microglobulin as a Measure of Glomerular Filtration RateActa Medica Scandinavica, 1985
- The blood serum concentration of cystatin C (γ-trace) as a measure of the glomerular filtration rateScandinavian Journal of Clinical and Laboratory Investigation, 1985
- Quantitation of γ-trace in human biological fluids: Indications for production in the central nervous systemScandinavian Journal of Clinical and Laboratory Investigation, 1979
- Prediction of Creatinine Clearance from Serum CreatinineNephron, 1976