Abstract
A modified couple stress theory is used to analyze the vibration frequency and sensitivity of an atomic force microscope (AFM) cantilever with a crack taking into account the normal interaction force between the cantilever tip and the sample surface. An explicit expression for the sensitivity of vibration modes of the cracked cantilever has been obtained using the relationship between the resonant frequency and tip-sample interaction. The results show that the modal sensitivity of the cantilever using the modified couple stress theory is smaller than that using the classical beam theory for the lower contact stiffness. However, the situation is reversed when the contact stiffness becomes larger. Furthermore, the effects of crack parameter, crack location and contact stiffness on the modal sensitivity of the cracked AFM cantilever are also investigated.