An In Vitro Study to Assess Determinant Features Associated With Fluid Sealing in the Design of Endotracheal Tube Cuffs and Exerted Tracheal Pressures*
Authors:
- Otavio Tavares RanzaniOROtavio Tavares Ranzani
- Joan Daniel MartiJMJoan Daniel Marti
- Valeria GiuntaVGValeria Giunta
- Nestor LuqueNLNestor Luque
- Valentina IsettaVIValentina Isetta
- Miquel FerrerMFMiquel Ferrer
- Ramon FarreRFRamon Farre
- Guilherme Leite PimentelGPGuilherme Leite Pimentel
- Antoni TorresATAntoni Torres
- 1 February 2013
- journal article
- research article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 41 (2) , 518-526
Abstract
To assess the structural characteristics involved in the design of high-volume low-pressure endotracheal tube cuffs that are associated with fluid sealing effectiveness and to determine the extent of transmitted tracheal pressures upon cuff inflation. In vitro study. Pneumology laboratories. Eight high-volume low-pressure cuffs of cylindrical or tapered shape, made of polyvinylchloride or polyurethane, were studied. Cuffs were tested within a tracheal model, oriented 30° above horizontal to assess 1 hr leakage of oropharyngeal secretions simulant at cuff internal pressures of 15-30 cm H2O. The four best performing cuffs were evaluated for 24 hrs using an internal pressure of 30 cm H2O. The extent of transmitted tracheal wall pressure throughout the cuff-trachea contact area was determined using an internal pressure sensor within a tracheal model upon cuff inflation up to 30 cm H2O. Outer diameter, length, and compliance of each cuff were assessed. Multivariate regression analysis was performed to identify the main determinants of simulant leakage rate. The cuff-trachea contact area and the percentage of tracheal wall pressure measurements greater than 50 cm H2O were computed. Cuff design characteristics significantly differ among tubes. The cuffs made of polyurethane showed the best short- and long-term sealing efficacy. Nevertheless, in the multivariate analysis, the cuff outer diameter (n: 288, p = 0.003) and length (n: 288, p < 0.001), along with the internal pressure (n: 288, p < 0.001), were the only predictors of simulant leakage rate. The tapered cuffs showed the lowest tracheal wall contact area (n: 96, p < 0.001). The tracheal wall pressure distribution pattern was heterogeneous, and the percentage of high tracheal wall pressure significantly differs among the cuffs (n: 96, p < 0.001). The high-volume low-pressure cuffs' outer diameter, length, material, and internal pressure are the main determinants of sealing efficacy. Despite internal pressure within the safe range, transmitted tracheal pressure is extremely heterogeneous and differs among cuffs, occasionally reaching localized, very high, unsafe levels.Keywords
This publication has 17 references indexed in Scilit:
- Comparison of a Novel Lycra Endotracheal Tube Cuff to Standard Polyvinyl Chloride Cuff and Polyurethane Cuff for Fluid Leak PreventionRespiratory Care, 2011
- Mechanical influences on fluid leakage past the tracheal tube cuff in a benchtop modelIntensive Care Medicine, 2011
- Variables affecting leakage past endotracheal tube cuffs: a bench studyIntensive Care Medicine, 2010
- Positive-end expiratory pressure reduces incidence of ventilator-associated pneumonia in nonhypoxemic patients*Critical Care Medicine, 2008
- Measurement of tracheal wall pressure: a comparison of three different in vitro techniquesAnaesthesia, 2008
- Effect of positive expiratory pressure and type of tracheal cuff on the incidence of aspiration in mechanically ventilated patients in an intensive care unit*Critical Care Medicine, 2008
- Outer diameter and shape of paediatric tracheal tube cuffs at higher inflation pressuresAnaesthesia, 2005
- Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tubeIntensive Care Medicine, 2003
- Performance Characteristics of Tracheal CuffsInternational Anesthesiology Clinics, 1974
- Evaluation of tracheal tube cuff designsCritical Care Medicine, 1973