Electromagnetic Interference by Transcutaneous Neuromuscular Electrical Stimulation in Patients with Bipolar Sensing Implantable Cardioverter Defibrillators:
- 1 February 2003
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 26 (2p1) , 626-629
Abstract
CREVENNA, R., et al.: Electromagnetic Interference by Transcutaneous Neuromuscular Electrical Stimulation in Patients with Bipolar Sensing Implantable Cardioverter Defibrillators: A Pilot Safety Study. Neuromuscular electrical stimulation (NMES) is a frequently applied therapy for the treatment of pain and a therapeutic option to increase thigh muscle strength and endurance capacity in patients with heart failure. Electromagnetic interference (EMI) by the signals with sensing of ICDs is possible. Eight patients with subpectoral ICD systems and different transvenous bipolar sensing leads were subjected to electrical stimulation of the neck and shoulder and of the thighs using different stimulation algorithms. EMI with ventricular sensing was detectable in three of eight subjects. EMI occurred during stimulation of the neck (n = 2) and thigh (n = 2). EMI by NMES with atrial sensing was seen in two of four subjects with dual chamber ICDs. The safety of peripheral NMES has to be individually tested as EMI can also occur in ICD patients with bipolar sensing. (PACE 2003; 26[Pt. I]:626–629)Keywords
This publication has 14 references indexed in Scilit:
- Washing Machine Associated 50 Hz Detected As Ventricular Fibrillation by An Implanted Cardioverter DefibrillatorPacing and Clinical Electrophysiology, 2001
- Incidence of Electromagnetic Interference in Implantable Cardioverter DefibrillatorsPacing and Clinical Electrophysiology, 2001
- Interactions Between Electronic Article Surveillance Systems and Implantable Cardioverter-DefibrillatorsCirculation, 1999
- Strength Improvement of Knee Extensor Muscles in Patients with Chronic Heart Failure by Neuromuscular Electrical StimulationArtificial Organs, 1999
- To the EditorPacing and Clinical Electrophysiology, 1999
- Electromagnetic Interference from a Muscle Stimulation Device Causing Discharge of an Implantable Cardioverter Defibrillator: Epicardial Bipolar and Endocardial Bipolar Sensing Circuits Are ComparedPacing and Clinical Electrophysiology, 1998
- Electromagnetic Interference and ICD Discharge Related to Chiropractic TreatmentPacing and Clinical Electrophysiology, 1998
- Inappropriate Shocks Delivered by an ICD as a Result of Sensed Potentials from a Transcutaneous Electronic Nerve Stimulation UnitPacing and Clinical Electrophysiology, 1998
- Effects of Low-Frequency Electrical Stimulation of Quadriceps and Calf Muscles in Patients With Chronic Heart FailureJournal of Cardiopulmonary Rehabilitation, 1998
- Erroneous Discharge of an Implantable Cardioverter Defibrillator Caused by an Electric RazorPacing and Clinical Electrophysiology, 1995