SOLID STATE IRIDIUM OXIDE–TITANIUM BASED SENSOR FOR FLOW INJECTION pH MEASUREMENTS
Authors:
- Saad S. M. HassanSHSaad S. M. Hassan
- S. A. M. MarzoukSMS. A. M. Marzouk
- N. M. BadawyNBN. M. Badawy
- 8 May 2002
- journal article
- Published by Taylor & Francis in Analytical Letters
- Vol. 35 (8) , 1301-1311
Abstract
Iridium oxide-coated titanium wire is prepared by controlled electrodeposition (current density 5 mA/cm2 for 10 min) and used as a solid state pH sensor. The sensor is sensitive for pH variations over the range 2–10 with calibration slopes of ˜ 65.5 and 59.1 mV/pH unit under static and hydrodynamic modes of operation, respectively. The general performance characteristics of the sensor are similar to those of the glass electrode for direct pH monitoring of solutions. The sensor exhibits reasonable stability (±0.005 pH), reproducibility (±2%) and fast response time (20–30 s). The method of preparation is simple, utilizes an inexpensive titanium metal substrate and offers the possibility of fabrication of sensors with various geometry and/or of micro and ultramicro sizes usable for flow injection analysis. No interferences are caused by alkali and alkaline earth metal cations and various anions at level ≥ 104 molar excess over H+.Keywords
This publication has 24 references indexed in Scilit:
- Solid-State Planar Microsensors for Selective Potentiometric Determination of EthylmorphineAnalytical Letters, 1999
- Interfacial properties of oxides with technological impact in electrochemistryAdvances in Colloid and Interface Science, 1996
- A solid-state pH sensor based on a Nafion-coated iridium oxide indicator electrode and a polymer-based silver chloride reference electrodeSensors and Actuators B: Chemical, 1994
- In Situ Electrochemical Sensor for Measurement in Nonconductive LiquidsJournal of the Electrochemical Society, 1993
- Metal-Metal Oxide and Metal Oxide Electrodes as pH SensorsCritical Reviews in Analytical Chemistry, 1989
- Evaluation of iridium oxide electrodes formed by potential cycling as pH probesThe Analyst, 1988
- Polycrystalline and monocrystalline antimony, iridium and palladium as electrode material for pH-sensing electrodesTalanta, 1986