Rare early prosthesis obstruction after mitral valve replacement: a case report and literature review
Open Access
- 2 July 2012
- journal article
- case report
- Published by Springer Nature in Journal of Cardiothoracic Surgery
- Vol. 7 (1) , 64
Abstract
As a dreadful complication after the mechanical heart valve replacement, prosthetic valve obstruction caused by pannus formation occurs increasingly with time. The authors here present a case of 42-year-old woman who was urgently admitted to hospital with acute heart failure symptoms due to the mechanical mitral valve failure only 3 months after surgery. Transthoracic and transesophageal echocardiography demonstrated that the bileaflet of the mitral prosthesis were completely immobilized with only a small transvalvular jet remained. During the reoperation, the reason of the prosthetic valve obstruction was attributed to the noncircular pannus ingrowth extending from the atrioventricular side. For a better understanding of the prosthetic valve dysfunction caused by pannus formation, the authors then compile a literature review to briefly discuss the status quo of the clinical characteristics of this uncommon complication.Keywords
This publication has 20 references indexed in Scilit:
- Transaortic chordae and pannus removal without redo mitral valve replacement in prosthetic mitral valve malfunctionEuropean Journal of Cardio-Thoracic Surgery, 2011
- St. Jude prosthetic valve obstruction in the mitral positionGeneral Thoracic and Cardiovascular Surgery, 2011
- Obstruction of St. Jude Medical Valves in the Aortic Position: Plasma Transforming Growth Factor Type Beta 1 in Patients With Pannus OvergrowthArtificial Organs, 2010
- Acute Failure of a St. Jude's Prosthetic Aortic Valve: Large Pannus Formation Masked by a Small ThrombusJournal of the American Society of Echocardiography, 2009
- Early failure of a mechanical bileaflet aortic valve prosthesis due to pannus: A rare complicationThe Journal of Thoracic and Cardiovascular Surgery, 2008
- Pannus formation on St. Jude Medical prosthetic aortic valve 23 years after initial operationEuropean Journal of Cardio-Thoracic Surgery, 2008
- Prosthetic heart valves: Types and echocardiographic evaluationInternational Journal of Cardiology, 2007
- Prevalence of pannus formation after aortic valve replacement: clinical aspects and surgical managementJournal of Artificial Organs, 2006
- Prosthetic valve thrombosis: twenty-year experience at the montreal heart instituteThe Journal of Thoracic and Cardiovascular Surgery, 2004
- Obstruction of St Jude medical valves in the aortic position: a consideration for pathogenic mechanism of prosthetic valve obstructionCardiovascular Surgery, 2002