Microfluidic routing of aqueous and organic flows at high pressures: fabrication and characterization of integrated polymer microvalve elements
Authors:
- 14 December 2004
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 5 (2) , 184-190
Abstract
This paper presents the first systematic engineering study of the impact of chemical formulation and surface functionalization on the performace of free-standing microfluidic polymer elements used for high-pressure fluid control in glass microsystems. System design, chemical wet-etch processes, and laser-induced polymerization techniques are described, and parametric studies illustrate the effects of polymer formulation, glass surface modification, and geometric constraints on system performance parameters. In particular, this study shows that highly crosslinked and fluorinated polymers can overcome deficiencies in previously-reported microvalve architectures, particularly limited solvent compatibility. Substrate surface modification is shown effective in reducing the friction of the polymer–glass interface and thereby facilitating valve actuation. A microchip one-way valve constructed using this architecture shows a 2 × 108 ratio of forward and backward flow rates at 7 MPa. This valve architecture is integrated on chip with minimal dead volumes (70 pl), and should be applicable to systems (including chromatography and chemical synthesis devices) requiring high pressures and solvents of varying polarity.Keywords
This publication has 19 references indexed in Scilit:
- On-Chip High-Pressure Picoliter Injector for Pressure-Driven Flow through Porous MediaAnalytical Chemistry, 2004
- A silicon-on-insulator based micro check valveJournal of Micromechanics and Microengineering, 2003
- A microfabricated suspended-tube chemical reactor for thermally efficient fuel processingJournal of Microelectromechanical Systems, 2003
- Solvent Compatibility of Poly(dimethylsiloxane)-Based Microfluidic DevicesAnalytical Chemistry, 2003
- A high-flow thermopneumatic microvalve with improved efficiency and integrated state sensingJournal of Microelectromechanical Systems, 2003
- Design and fabrication of microfluidic devices for multiphase mixing and reactionJournal of Microelectromechanical Systems, 2002
- Surface micromachined paraffin-actuated microvalveJournal of Microelectromechanical Systems, 2002
- Microfluidic Large-Scale IntegrationScience, 2002
- Fabrication and characterization of hydrogel-based microvalvesJournal of Microelectromechanical Systems, 2002
- Monolithic Microfabricated Valves and Pumps by Multilayer Soft LithographyScience, 2000