Evaluation of an Atlas-Based PET Head Attenuation Correction Using PET/CT & MR Patient Data
- 20 August 2013
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 60 (5) , 3383-3390
Abstract
The goal of this study was to compare MR-based PET patient attenuation correction (AC) to CT-based AC in the head using clinical whole-body FDG-PET patient data obtained from a tri-modality PET/CT & MR setup. The MR-based AC utilizes an atlas-based approach, registering the patient's MR images to a CT-based atlas, producing `pseudoCT' images. Thirteen clinical whole-body FDG patients were included in this study. PET mean activity concentration values were measured and compared in six ~ 15 ml volumes-of-interest throughout the brain tissue. The AC methods compared to CT-based AC were segmentation of the CT (air, fat, soft tissue) and atlas-based MR-AC. Results: PET activity concentration was systematically under-estimated on average by 1.32 kBq/ml (4.9%) when using the segmented CT-based AC, mainly due to lack of attenuation correction for skull bone. Using the atlas-based method, the error was reduced to 0.03 kBq/ml (0.2%) on average. PET image visualization demonstrated spatial variations in activity concentration accuracy induced by the AC methods that were consistent with the approximations in each method. Conclusion: The results demonstrate that the atlas-based AC in the head provides adequate PET quantitation and image quality as compared to methods that do not account for bone.Keywords
This publication has 24 references indexed in Scilit:
- Errors in MR-based attenuation correction for brain imaging with PET/MR scannersNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- MR-Based PET Attenuation Correction for PET/MR ImagingSeminars in Nuclear Medicine, 2013
- PET–MR imaging using a tri-modality PET/CT–MR system with a dedicated shuttle in clinical routineMagnetic Resonance Materials in Physics, Biology and Medicine, 2012
- The effect of errors in segmented attenuation maps on PET quantificationMedical Physics, 2011
- Attenuation Correction Methods Suitable for Brain Imaging with a PET/MRI Scanner: A Comparison of Tissue Atlas and Template Attenuation Map ApproachesJournal of Nuclear Medicine, 2011
- MR‐based attenuation correction for hybrid PET‐MR brain imaging systems using deformable image registrationMedical Physics, 2010
- MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas RegistrationJournal of Nuclear Medicine, 2008
- Template based attenuation correction for PET in MR-PET scanners2008
- Atlas-guided non-uniform attenuation correction in cerebral 3D PET imagingNeuroImage, 2005
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994