The first prospective implementation of markerless lung target tracking in an experimental quality assurance procedure on a standard linear accelerator
- 1 January 2020
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 65 (2) , 025008
Abstract
The ability to track tumour motion without implanted markers on a standard linear accelerator (linac) could enable wide access to real-time adaptive radiotherapy for cancer patients. We previously have retrospectively validated a method for 3D markerless target tracking using intra-fractional kilovoltage (kV) projections acquired on a standard linac. This paper presents the first prospective implementation of markerless lung target tracking on a standard linac and its quality assurance (QA) procedure. The workflow and the algorithm developed to track the 3D target position during volumetric modulated arc therapy treatment delivery were optimised. The linac was operated in clinical QA mode, while kV projections were streamed to a dedicated computer using a frame-grabber software. The markerless target tracking accuracy and precision were measured in a lung phantom experiment under the following conditions: static localisation of seven distinct positions, dynamic localisation of five patient-measured motion traces, and dynamic localisation with treatment interruption. The QA guidelines were developed following the AAPM Task Group 147 report with the requirement that the tracking margin components, the margins required to account for tracking errors, did not exceed 5 mm in any direction. The mean tracking error ranged from 0.0 to 0.9 mm (left–right), −0.6 to −0.1 mm (superior–inferior) and −0.7 to 0.1 mm (anterior–posterior) over the three tests. Larger errors were found in cases with large left–right or anterior–posterior and small superior–inferior motion. The tracking margin components did not exceed 5 mm in any direction and ranged from 0.4 to 3.2 mm (left–right), 0.7 to 1.6 mm (superior–inferior) and 0.8 to 1.5 mm (anterior–posterior). This study presents the first prospective implementation of markerless lung target tracking on a standard linac and provides a QA procedure for its safe clinical implementation, potentially enabling real-time adaptive radiotherapy for a large population of lung cancer patients.Keywords
Funding Information
- National Health and Medical Research Council (Early Career Fellowship APP1120333, Senior Principal Research Fellowshi)
- Cancer Institute NSW (Career Development Fellowship, Early Career Fellowship CS00481)
This publication has 53 references indexed in Scilit:
- Markerless EPID image guided dynamic multi-leaf collimator tracking for lung tumorsPhysics in Medicine & Biology, 2013
- Monitoring tumor motion by real time 2D/3D registration during radiotherapyRadiotherapy and Oncology, 2012
- Stereotactic body radiation therapy: The report of AAPM Task Group 101Medical Physics, 2010
- Marker-free lung tumor trajectory estimation from a cone beam CT sinogramPhysics in Medicine & Biology, 2010
- The diaphragm as an anatomic surrogate for lung tumor motionPhysics in Medicine & Biology, 2009
- Synchronized tumour tracking with electromagnetic transponders and kV x-ray imaging: evaluation based on a thorax phantomPhysics in Medicine & Biology, 2008
- An analysis of thoracic and abdominal tumour motion for stereotactic body radiotherapy patientsPhysics in Medicine & Biology, 2008
- Feasibility of using ultrasound for real‐time tracking during radiotherapyMedical Physics, 2005
- Errors and margins in radiotherapySeminars in Radiation Oncology, 2004
- The Cyberknife: A Frameless Robotic System for RadiosurgeryStereotactic and Functional Neurosurgery, 1997