Abstract
Vitamin K antagonists, such as warfarin, have a narrow therapeutic window; patients on these therapies therefore require regular international normalized ratio (INR) monitoring to maintain optimal dosing. This involves periodic checks and laboratory testing using venepuncture, which are often perceived as a burden. This study aimed to determine the accuracy and precision of the LumiraDx INR Test, a new point-of-care in vitro diagnostic platform, in an anticoagulation clinic setting. In this observational, cross-sectional study, precision of the LumiraDx INR Test was assessed using paired replicate samples (n = 366) and 3 test strip lots. Accuracy was determined by comparing capillary blood INR, ascertained by the LumiraDx INR Test, with venous plasma INR, measured by the laboratory reference instrument, the IL ACL ELITE Pro. Furthermore, INR was assessed across a range of hematocrit (25%-55%). In addition, feedback was collected from health-care professionals via a self-completed questionnaire. This trial was registered at ClinicalTrials.gov (NCT03682419). The precision (% coefficient of variation) of the LumiraDx INR Test was <4 when samples were applied by direct application or via a capillary transfer pipette, as well as between test strip lots. Accuracy of the LumiraDx INR Test, across the INR range of 0.8 to 7.5, was confirmed by a strong correlation of 0.965 (95% confidence interval: 0.959-0.970) when compared with the IL ACL ELITE Pro, which was maintained across the hematocrit range. Feedback from health-care professionals indicated that the instructions given by the system were easy to follow. In conclusion, the strong agreement between the LumiraDx Platform INR point-of-care test and the IL ACL ELITE Pro laboratory reference system, as well as between the different application methods and test lots, indicates that it can provide a rapid, accurate, and reliable INR analysis.Keywords
Funding Information
- LumiraDx UK Ltd
This publication has 14 references indexed in Scilit:
- American Society of Hematology 2018 guidelines for management of venous thromboembolism: optimal management of anticoagulation therapyBlood Advances, 2018
- Benefits and Pitfalls of Point-of-Care Coagulation Testing for Anticoagulation ManagementAmerican Journal of Clinical Pathology, 2018
- Warfarin: The End or the End of One Size Fits All Therapy?Journal of Personalized Medicine, 2018
- Antithrombotic therapy in atrial fibrillation associated with valvular heart disease: a joint consensus document from the European Heart Rhythm Association (EHRA) and European Society of Cardiology Working Group on Thrombosis, endorsed by the ESC Working Group on Valvular Heart Disease, Cardiac Arrhythmia Society of Southern Africa (CASSA), Heart Rhythm Society (HRS), Asia Pacific Heart Rhythm Society (APHRS), South African Heart (SA Heart) Association and Sociedad Latinoamericana de Estimulación Cardíaca y Electrofisiología (SOLEACE)EP Europace, 2017
- Time to Revisit the Time in the Therapeutic RangeJournal of Atrial Fibrillation, 2017
- The optimal management of anti-thrombotic therapy after valve replacement: certainties and uncertaintiesEuropean Heart Journal, 2014
- Precision and accuracy of point‐of‐care testing coagulometers used for self‐testing and self‐management of oral anticoagulation therapyJournal of Thrombosis and Haemostasis, 2012
- Guidelines on oral anticoagulation with warfarin – fourth editionBritish Journal of Haematology, 2011
- Patient satisfaction with point-of-care testing in general practiceBritish Journal of General Practice, 2010
- International Normalized Ratios (INR): the first 20 yearsJournal of Thrombosis and Haemostasis, 2004