A randomised trial evaluating mask ventilation using electrical impedance tomography during anesthetic induction: one-handed technique versus two-handed technique
Open Access
- 28 June 2022
- journal article
- research article
- Published by IOP Publishing in Physiological Measurement
- Vol. 43 (6) , 064004
Abstract
Objective. Mask positive-pressure ventilation could lead to lung ventilation inhomogeneity, potentially inducing lung function impairments, when compared with spontaneous breathing. Lung ventilation inhomogeneity can be monitored by chest electrical impedance tomography (EIT), which could increase our understanding of mask ventilation-derived respiratory mechanics. We hypothesized that the two-handed mask holding ventilation technique resulted in better lung ventilation, reflected by respiratory mechanics, when compared with the one-handed mask holding technique. Approach. Elective surgical patients with healthy lungs were randomly assigned to receive either one-handed mask holding (one-handed group) or two-handed mask holding (two-handed group) ventilation. Mask ventilation was performed by certified registered anesthesiologists, during which the patients were mechanically ventilated using the pressure-controlled mode. EIT was used to assess respiratory mechanics, including ventilation distribution, global and regional respiratory system compliance ( C RS ), expiratory tidal volume (TVe) and minute ventilation volume. Hemodynamic parameters and the PaO 2 -FiO 2 ratio were also recorded. Main results. Eighty adult patients were included in this study. Compared with spontaneous ventilation, mask positive-pressure ventilation caused lung ventilation inhomogeneity with both one-handed(global inhomogeneity index: 0.40 ± 0.07 versus 0.50 ± 0.15; P < 0.001) and two-handed mask holding (0.40 ± 0.08 versus 0.50 ± 0.13; P < 0.001). There were no differences in the global inhomogeneity index ( P = 0.948) between the one-handed and two-handed mask holding. Compared with the one-handed mask holding, the two-handed mask holding was associated with higher TVe (552.6 ± 184.2 ml versus 672.9 ± 156.6 ml, P = 0.002) and higher global C RS (46.5 ± 16.4 ml/cmH 2 O versus 53.5 ± 14.5 ml/cmH 2 O, P = 0.049). No difference in PaO 2 -FiO 2 ratio was found between both holding techniques ( P = 0.743). Significance. The two-handed mask holding technique could not improve the inhomogeneity of lung ventilation when monitored by EIT during mask ventilation although it obtained larger expiratory tidal volumes than the one-handed mask holding technique.Keywords
Funding Information
- Science and Technology Commission of Shanghai Municipality (20Y11906200)
This publication has 25 references indexed in Scilit:
- Objective description of mask ventilationBritish Journal of Anaesthesia, 2016
- Noninvasive ventilation during spontaneous breathing anesthesia: an observational study using electrical impedance tomographyJournal of Clinical Anesthesia, 2016
- Perioperative assessment of regional ventilation during changing body positions and ventilation conditions by electrical impedance tomographyBritish Journal of Anaesthesia, 2016
- Induction of general anaesthesia by effect‐site target‐controlled infusion of propofol: influence of pharmacokinetic model and k e0 valueAnaesthesia, 2014
- Effect of PEEP and Tidal Volume on Ventilation Distribution and End-Expiratory Lung Volume: A Prospective Experimental Animal and Pilot Clinical StudyPLOS ONE, 2013
- Spontaneous Breathing during General Anesthesia Prevents the Ventral Redistribution of Ventilation as Detected by Electrical Impedance TomographyAnesthesiology, 2012
- A Two-handed Jaw-thrust Technique Is Superior to the One-handed “EC-clamp” Technique for Mask Ventilation in the Apneic Unconscious PersonAnesthesiology, 2010
- Spatial distribution of sequential ventilation during mechanical ventilation of the uninjured lung: an argument for cyclical airway collapse and expansionBMC Pulmonary Medicine, 2010
- End-expiratory lung impedance change enables bedside monitoring of end-expiratory lung volume changeIntensive Care Medicine, 2003
- Airway closure, atelectasis and gas exchange during general anaesthesiaBritish Journal of Anaesthesia, 1998