Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience
Authors:
- 4 September 2009
- journal article
- cardiac
- Published by Springer Nature in European Radiology
- Vol. 19 (11) , 2576-2583
Abstract
Objective We evaluated radiation exposure and image quality of a new coronary CT angiography protocol, high-pitch spiral acquisition, using dual source CT (DSCT). Material and methods Coronary CTA was performed in 25 consecutive patients with a stable heart rate of 60 bpm or less after premedication, using 2 × 128 0.6-mm sections, 38.4-mm collimation width and 0.28-s rotation time. Tube settings were 100 kV/320 mAs and 120 kV/400 mAs for patients below and above 100-kg weight, respectively. Data acquisition was prospectively ECG-triggered at 60% of the R–R interval using a pitch of 3.2 (3.4 for the last 10 patients). Images were reconstructed with 75-ms temporal resolution, 0.6-mm slice thickness and 0.3-mm increment. Image quality was evaluated using a four-point scale (1 = excellent, 4 = unevaluable). Results Mean range of data acquisition was 113 ± 22 mm, mean duration was 268 ± 23 ms. Of 363 coronary artery segments, 327 had an image quality score of 1, and only 2 segments were rated as “unevaluable”. Mean dose–length product (DLP) was 71 ± 23 mGy cm, mean effective dose was 1.0 ± 0.3 mSv (range 0.78–2.1 mSv). For 21 patients with a body weight below 100 kg, mean DLP was 63 ± 5 mGy cm (0.88 ± 0.07 mSv; range 0.78–0.97 mSv). Conclusion Prospectively ECG-triggered high-pitch spiral CT acquisition provides high and stable image quality at very low radiation dose.Keywords
This publication has 26 references indexed in Scilit:
- Image Quality in a Low Radiation Exposure Protocol for Retrospectively ECG-Gated Coronary CT AngiographyAmerican Journal of Roentgenology, 2009
- Application‐ and patient size‐dependent optimization of x‐ray spectra for CTMedical Physics, 2009
- Algorithm for radiation dose reduction with helical dual source coronary computed tomography angiography in clinical practiceJournal of Cardiovascular Computed Tomography, 2008
- Tips to minimize radiation exposureJournal of Cardiovascular Computed Tomography, 2008
- Prospective versus Retrospective ECG Gating for 64-Detector CT of the Coronary Arteries: Comparison of Image Quality and Patient Radiation DoseRadiology, 2008
- Prospective versus Retrospective ECG-gated 64-Detector Coronary CT Angiography: Assessment of Image Quality, Stenosis, and Radiation DoseRadiology, 2008
- Prospectively Gated Transverse Coronary CT Angiography versus Retrospectively Gated Helical Technique: Improved Image Quality and Reduced Radiation DoseRadiology, 2008
- Estimating Risk of Cancer Associated With Radiation Exposure From 64-Slice Computed Tomography Coronary AngiographyJAMA, 2007
- Strategies for reduction of radiation dose in cardiac multislice CTEuropean Radiology, 2007
- Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulationEuropean Radiology, 2002