CD4 Lymphocyte Enumeration and Hemoglobin Assessment Aid for Priority Decisions: A Multisite Evaluation of the BD FACSPresto™ System
Authors:
Open Access
- 24 October 2017
- journal article
- Published by Bentham Science Publishers Ltd. in The Open AIDS Journal
- Vol. 11 (1) , 76-90
Abstract
Background: : The BD FACSPresto™system uses capillary and venous blood to measure CD4 absolute counts (CD4), %CD4 in lymphocytes, and hemoglobin (Hb) in approximately 25 minutes. CD4 cell count is used with portable CD4 counters in resource-limited settings to manage HIV/AIDS patients. A method comparison was performed using capillary and venous samples from seven clinical laboratories in five countries. The BD FACSPresto system was assessed for variability between laboratory, instrument/operators, cartridge lots and within-run at four sites.Methods: : Samples were collected under approved voluntary consent. EDTA-anticoagulated venous samples were tested for CD4 and %CD4 T cells using the gold-standard BD FACSCalibur™system, and for Hb, using the Sysmex®KX-21N™analyzer. Venous and capillary samples were tested on the BD FACSPresto system. Matched data was analyzed for bias (Deming linear regression and Bland-Altman methods), and for concordance around the clinical decision point. The coefficient of variation was estimated per site, instrument/operator, cartridge-lot and between-runs.Results: : For method comparison, 93% of the 720 samples were from HIV-positive and 7% from HIV-negative or normal subjects. CD4 and %CD4 T cells venous and capillary results gave slopes within 0.96–1.05 and R2≥0.96; Hb slopes were ≥1.00 and R2≥0.89. Variability across sites/operators gave %CV <5.8% for CD4 counts, <1.9% for %CD4 and <3.2% for Hb. The total %CV was <7.7% across instrument/cartridge lot.Conclusion: : The BD FACSPresto system provides accurate, reliable, precise CD4/%CD4/Hb results compared to gold-standard methods, irrespective of venous or capillary blood sampling. The data showed good agreement between the BD FACSPresto, BD FACSCalibur and Sysmex systems.Keywords
This publication has 36 references indexed in Scilit:
- Trends in mortality and loss to follow-up in HIV care at the Nkongsamba Regional hospital, CameroonBMC Research Notes, 2013
- Maternal anaemia and duration of zidovudine in antiretroviral regimens for preventing mother-to-child transmission: a randomized trial in three African countriesBMC Infectious Diseases, 2013
- Biological variability of lymphocyte subsets of human adults' bloodClinica Chimica Acta, 2013
- Performance evaluation of the Pima™ point‐of‐care CD4 analyser using capillary blood sampling in field tests in South AfricaJournal of the International AIDS Society, 2012
- Utility of the point of care CD4 analyzer, PIMA, to enumerate CD4 counts in the field settings in IndiaAIDS Research and Therapy, 2012
- Absolute and percent CD4+ T-cell enumeration by flow cytometry using capillary bloodJournal of Immunological Methods, 2011
- Establishment of reference CD4+ T cell values for adult Indian populationAIDS Research and Therapy, 2011
- Use of capillary blood count parameters in adultsVox Sanguinis, 2007
- The impact of malnutrition on survival and the CD4 count response in HIV‐infected patients starting antiretroviral therapyHIV Medicine, 2006
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986