Abstract
Fiber-optic reflectance measurements were performed on scattering phantoms containing red blood cells. The oxygen pressure in the phantom was decreased by nitrogen gas bubbling, while carefully controlling and monitoring the temperature and pH (7.4). The blood oxygen saturation was extracted from the optical spectroscopic measurements by fitting the spectra over the wavelength range of to a model that includes the effects of pigment packaging. The effect of using different oxy- and deoxyhemoglobin extinction spectra on the quality and accuracy of the fits is analyzed. The optically extracted is compared with the standard oxygen dissociation curve (ODC). Depending on the choice of hemoglobin extinction spectra the absolute deviation between the ODC and the optically extracted is less than 2.5% over the measured range of saturations (4%–99%).
Keywords
Funding Information
- Dutch Technology Foundation (STW)
- The Netherlands Foundation for Scientific Research (NWO)
- Technology Program of the Ministry of Economic Affairs
This publication has 12 references indexed in Scilit:
- Confidence intervals on fit parameters derived from optical reflectance spectroscopy measurementsJournal of Biomedical Optics, 2008
- Non-invasive measurement of the morphology and physiology of oral mucosa by use of optical spectroscopyOral Oncology, 2008
- Empirical model description of photon path length for differential path length spectroscopy: combined effect of scattering and absorptionJournal of Biomedical Optics, 2008
- Analytical model of light reflectance for extraction of the optical properties in small volumes of turbid mediaApplied Optics, 2007
- Monte Carlo-based inverse model for calculating tissue optical properties Part I: Theory and validation on synthetic phantomsApplied Optics, 2006
- Determination of optical properties of human blood in the spectral range 250 to 1100 nm using Monte Carlo simulations with hematocrit-dependent effective scattering phase functionsJournal of Biomedical Optics, 2006
- Hemoglobin oxygen saturations in phantoms and in vivo from measurements of steady‐state diffuse reflectance at a single, short source‐detector separationMedical Physics, 2004
- In vivo measurement of the local optical properties of tissue by use of differential path-length spectroscopyOptics Letters, 2004
- A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivoMedical Physics, 1992
- Light scattering in lntralipid-10% in the wavelength range of 400–1100 nmApplied Optics, 1991