Graphical and Numerical Evaluation of Continuous Glucose Sensing Time Lag
- 1 March 2009
- journal article
- research article
- Published by SAGE Publications in Diabetes Technology & Therapeutics
- Vol. 11 (3) , 139-143
Abstract
Aim: This study introduces a new method for graphical and numerical evaluation of time lags typically associated with subcutaneous glucose sensing, based on Poincaré-type plot and a maximum statistical agreement criterion. Methods: The proposed method is illustrated by retrospective analysis of 56 continuous glucose monitor (CGM) time series collected by the FreeStyle Navigator™ (Abbott Diabetes Care, Alameda, CA) from 28 patients with type 1 diabetes mellitus, each wearing simultaneously two sensors (on arm and abdomen) and parallel reference blood glucose (BG) collected with a reference YSI (Yellow Springs, OH) analyzer every 15 min. The average duration of a time series was 111 h; there were approximately 10,000 sensor–reference data pairs. Results: When sliding in time CGM readings versus BG, the point of minimal spread of a Poincaré-type plot marks visually the time of CGM delay. The same point is numerically estimated by minimizing the distance between BG and CGM readings. The average observed time lag between reference BG and CGM was 12.5 min. Stratified by BG rate of change, the time lag was longer (16.8 min) when BG was falling, compared to steady or rising BG (11.7 min and 9.9 min, respectively) (P < 0.005). The time lags at the two sensor locations were not significantly different: 12.4 min on the arm, 12.6 min on the abdomen. Conclusions: In this data set, substantial blood-to-sensor time delays were observed, possibly because of both blood-to-interstitial glucose transport and instrumental delay. Analysis of BG-CGM co-dynamics that is free from mathematical approximation of glucose fluctuations resulted in convenient visualization and numerical estimation of these delays.Keywords
This publication has 23 references indexed in Scilit:
- Relationship Between Interstitial And Blood Glucose in Type 1 Diabetes Patients: Delay And The Push-Pull Phenomenon RevisitedDiabetes Technology & Therapeutics, 2007
- Enhancing Poincare plot information via sampling ratesApplied Mathematics and Computation, 2007
- Quantifying Temporal Glucose Variability in Diabetes via Continuous Glucose Monitoring: Mathematical Methods and Clinical ApplicationDiabetes Technology & Therapeutics, 2005
- Interstitial fluid glucose dynamics during insulin-induced hypoglycaemiaDiabetologia, 2005
- Heart rate variability in young patients with diabetes mellitus and healthy subjects explored by Poincaré and sequence plotsClinical Physiology and Functional Imaging, 2005
- A Novel Approach to Mitigating the Physiological Lag Between Blood and Interstitial Fluid Glucose MeasurementsDiabetes Technology & Therapeutics, 2004
- A Continuous Glucose Sensor Based on Wired Enzyme ™ Technology - Results from a 3-Day Trial in Patients with Type 1 DiabetesDiabetes Technology & Therapeutics, 2003
- Can Interstitial Glucose Assessment Replace Blood Glucose Measurements?Diabetes Technology & Therapeutics, 2000
- Fascinating rhythm: A primer on chaos theory and its application to cardiologyAmerican Heart Journal, 1990
- Detecting strange attractors in turbulence1981