Characterization of an Innovative p-type Epitaxial Diode for Dosimetry in Modern External Beam Radiotherapy
- 3 December 2013
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 60 (6) , 4705-4712
Abstract
Due to the ever-increasing complexity of treatment modalities in radiation therapy, there has been a greater need for detectors to perform quality assurance to ensure patients are treated correctly and safely. Modern radiation therapy techniques involve small field sizes, high dose gradients, and varying intensity of energy and rate. The ideal dosimeter for this treatment should display high spatial resolution, high linearity, accuracy, and radiation hardness. Silicon detectors have been widely used for radiotherapy measurements and have many attractive qualities as a dosimeter; weaknesses of silicon detectors are, however, decreases in sensitivity with accumulated dose. The Centre for Medical Radiation Physics has developed a new technology with an unusual charge collection efficiency variation with accumulated dose which stabilizes the response of the detector within ±5% after 120 kGy photon irradiation. The sensor has been also characterized by irradiation by an 18 MV medical LINAC with sensitivity to a photoneutron-induced damage of less than 0.5%/100 Gy. The radiation damage mechanism has been validated by TCAD simulations which confirmed the mechanism behind the CCE increase as a function of the accumulated dose.Keywords
This publication has 32 references indexed in Scilit:
- A review on photoneutrons characteristics in radiation therapy with high-energy photon beamsReports of Practical Oncology and Radiotherapy, 2010
- The effect of field modifier blocks on the fast photoneutron dose equivalent from two high-energy medical linear acceleratorsRadiation Protection Dosimetry, 2007
- Epitaxial silicon devices for dosimetry applicationsApplied Physics Letters, 2007
- Electronic dosimetry in radiation therapyRadiation Measurements, 2006
- The ideal dosimeter for intensity modulated radiation therapy (IMRT): What is required?Journal of Physics: Conference Series, 2004
- Low-temperature radiation-stimulated gettering of impurities and defects in silicon by layers of porous siliconTechnical Physics Letters, 1997
- General specifications for silicon semiconductors for use in radiation dosimetryPhysics in Medicine & Biology, 1987
- Evaluation of temperature effects in p-type silicon detectorsPhysics in Medicine & Biology, 1986
- Electric field effect on radiation-induced gettering of defects in n+-n-n++GaAsPhysica Status Solidi (a), 1984
- Effects of radiation damage on p-type silicon detectorsPhysics in Medicine & Biology, 1983