Model‐based prediction of portal dose images during patient treatment
Authors:
- 27 February 2013
- journal article
- Published by Wiley in Medical Physics
- Vol. 40 (3) , 031713
Abstract
Purpose:: Dosimetric verification of radiation therapy is crucial when delivering complex treatments like intensity modulated radiation therapy (IMRT) or volumetric modulated arc therapy (VMAT. Pretreatment verification, characterized by methods applied without the patient present and before the treatment start date, is typically carried out at most centers.In vivo dosimetric verification, characterized by methods applied with the patient present, is not commonly carried out in the clinic. This work presents a novel, model‐based EPID dosimetry method that could be used for routine clinical in vivo patient treatment verification.Methods:: The authors integrated a detailed fluence model with a patient scatter prediction model that uses a superposition of scatter energy fluence kernels, generated via Monte Carlo techniques, to determine patient scatter fluence delivered to the EPID. The total dose to the EPID was calculated using the sum of convolutions of the calculated energy fluence distribution entering the EPID with monoenergetic dose kernels, specific to the a‐Si EPID. Measured images with simple, square fields delivered to slab phantoms were validated against predicted images. Measured and predicted images acquired during the delivery of IMRT fields to slabs and an anthropomorphic phantom were compared using theχ‐comparison for 3% dose difference and 3 mm distance‐to‐agreement criteria.Results:: Predicted and measured images of the square fields with slabs in the field agreed within 2.5%. Predicted portal dose images of clinical IMRT fields delivered to slabs and an anthropomorphic phantom agreed with measured images within 3% and 3 mm for an average of at least 97% of the infield pixels (defined as >10% maximum field dose) for each case, over all fields.Conclusions:: This work presents the first validation of the integration of a comprehensive fluence model with a patient and EPID radiation transport model that accounts for patient transmission, including complex factors such as patient scatter and the energy response of the a‐Si detector. The portal dose image prediction model satisfies the 3% and 3 mm criteria for IMRT fields delivered to slab phantoms and could be used for patient treatment verification.Keywords
This publication has 45 references indexed in Scilit:
- Simplifying EPID dosimetry for IMRT treatment verificationMedical Physics, 2011
- Catching errors with in vivo EPID dosimetryMedical Physics, 2010
- 3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetryRadiotherapy and Oncology, 2010
- A literature review of electronic portal imaging for radiotherapy dosimetryRadiotherapy and Oncology, 2008
- The next step in patient-specific QA: 3D dose verification of conformal and intensity-modulated RT based on EPID dosimetry and Monte Carlo dose calculationsRadiotherapy and Oncology, 2008
- 3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CTRadiotherapy and Oncology, 2008
- Treatment verification in the presence of inhomogeneities using EPID‐based three‐dimensional dose reconstructionMedical Physics, 2007
- Replacing Pretreatment Verification With In Vivo EPID Dosimetry for Prostate IMRTInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Clinical experience with EPID dosimetry for prostate IMRT pre‐treatment dose verificationMedical Physics, 2006
- A Monte Carlo based three‐dimensional dose reconstruction method derived from portal dose imagesMedical Physics, 2006