Experimental determination ofkQfactors for cylindrical ionization chambers in 10 cm × 10 cm and 3 cm × 3 cm photon beams from 4 MV to 25 MV
Authors:
- R P KapschRKR P Kapsch
- 14 July 2014
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (15) , 4227-4246
Abstract
For the ionometric determination of absorbed dose to water, Dw, in megavoltage photon beams from a linear accelerator, beam-quality-dependent correction factors, kQ, are used for the ionization chambers. By using a water calorimeter, these factors can be determined experimentally and with substantially lower standard uncertainties compared to calculated values of the kQ, which are published in various dosimetry protocols. In this investigation, kQ for different types of cylindrical ionization chambers (NE 2561, NE 2571, FC 65 G) were determined experimentally in 10 cm × 10 cm photon beams from 4 MV to 25 MV (corresponding beam quality index TPR20,10 from 0.64 to 0.80). The measurements were carried out at the linear accelerator facility of the Physikalisch-Technische Bundesanstalt. It is shown that the kQ factors for a single ionization chamber in 10 cm × 10 cm photon beams can be measured with a relative standard uncertainty of 0.31%. In addition to these measurements in 10 cm × 10 cm fields, kQ factors for the NE 2561 chamber were also determined in smaller 3 cm × 3 cm photon beams between 6 MV and 25 MV. In this case, relative standard uncertainties between 0.35 % and 0.38 % are achieved for the kQ factors. It is found for this ionization chamber, that the ratio of the kQ factors in 3 cm × 3 cm and in 10 cm × 10 cm beams increases with increasing TPR20,10 to reach a value of 1.0095 at TPR20,10 = 0.8 with a relative standard uncertainty of 0.4 %.Keywords
This publication has 27 references indexed in Scilit:
- Monte Carlo calculated correction factors for diodes and ion chambers in small photon fieldsPhysics in Medicine & Biology, 2013
- Comparison of absorbed-dose-to-water units for Co-60 and high-energy x-rays between PTB and LNE–LNHBMetrologia, 2012
- Calculation of for several small detectors and for two linear accelerators using Monte Carlo simulationsMedical Physics, 2011
- A new formalism for reference dosimetry of small and nonstandard fieldsMedical Physics, 2008
- Calorimetric determination ofkQfactors for NE 2561 and NE 2571 ionization chambers in 5 cm × 5 cm and 10 cm × 10 cm radiotherapy beams of 8 MV and 16 MV photonsPhysics in Medicine & Biology, 2007
- The PTB water calorimeter for the absolute determination of absorbed dose to water in60Co radiationMetrologia, 2006
- Water calorimetry: A correction to the heat defect calculationsJournal of Research of the National Institute of Standards and Technology, 2002
- A comparison between calculated and experimentalkQphoton beam quality correction factorsPhysics in Medicine & Biology, 2000
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- The Cyberknife: A Frameless Robotic System for RadiosurgeryStereotactic and Functional Neurosurgery, 1997