High-flow nasal cannula as a device to provide continuous positive airway pressure in infants

Authors:

Abstract
To measure the intrapharyngeal pressure (IPP) generated by high-flow nasal cannula (HFNC) at varying flow rates up to 5 l min−1. We studied 14 infants on HFNC or NCPAP using IPP manometry to measure the IPP generated at flows of 1, 2, 3, 4 and 5 l min−1. On average flow rates of 1, 2, 3, 4 and 5 l min−1 generated an IPP (cm of H2O) of 1.70±0.34, 1.75±0.2, 2.62±0.28, 3.78±0.44 and 4.84±0.51 respectively. HFNC delivers significant IPP and is potentially a well-tolerated and viable option to provide CPAP at flows of 3 min−1 in infants with respiratory distress.