Coronary artery stents microfabricated from planar metal foil: design, fabrication, and mechanical testing
Authors:
- 8 October 2003
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 10846999,p. 462-465
Abstract
This paper presents a new approach to the design and manufacture of coronary artery stents, which permits the use of planar batch fabrication technologies. Stent samples with different wall patterns have been fabricated from 50-/spl mu/m thick stainless steel foil using micro-electro-discharge machining, and expanded to tubular shapes by using angioplasty balloons. The stents exhibit diameter variations of </spl plusmn/4 %, almost zero radial recoil after deflation of the balloon, and longitudinal shrinkage of 10/spl times/ higher.Keywords
This publication has 5 references indexed in Scilit:
- Laser microprocessing of stent for medical therapy2002
- Batch mode micro-electro-discharge machiningJournal of Microelectromechanical Systems, 2002
- Finite-element simulation of stent expansionJournal of Materials Processing Technology, 2002
- Radial force of coronary stents: A comparative analysisCatheterization and Cardiovascular Interventions, 1999
- Strength, Elasticity, and Plasticity of Expandable Metal Stents: In Vitro Studies with Three Types of StressJournal of Vascular and Interventional Radiology, 1994