CT Angiography of Neonates and Infants: Comparison of Radiation Dose and Image Quality of Target Mode Prospectively ECG-Gated 320-MDCT and Ungated Helical 64-MDCT
Authors:
- 1 February 2015
- journal article
- research article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 204 (2) , W184-W191
Abstract
OBJECTIVE. The purpose of this study was to evaluate the radiation dose and image quality of target mode prospectively ECG-gated volumetric CT angiography (CTA) performed with a 320-MDCT scanner compared with the radiation dose and image quality of ungated helical CTA performed with a 64-MDCT scanner. MATERIALS AND METHODS. An experience with CTA for cardiovascular indications in neonates and infants 0-6 months old was retrospectively assessed. Radiation doses and quantitative and qualitative image quality scores of 28 CTA examinations performed with a 320-MDCT scanner and volumetric target mode prospective ECG gating plus iterative reconstruction (target mode) were compared with the doses and scores of 28 CTA examinations performed with a 64-MDCT scanner and ungated helical scanning plus filtered back projection reconstruction (ungated mode). All target mode studies were performed during free breathing. Seven ungated CTA examinations (25%) were performed with general endotracheal anesthesia. The findings of 17 preoperative CTA examinations performed in target mode were also compared with surgical reports for evaluation of diagnostic accuracy. RESULTS. All studies performed with target mode technique were diagnostic for the main clinical indication. Effective doses were significantly lower in the target mode group (0.51 ± 0.19 mSv) compared with the ungated mode group (4.8 ± 1.4 mSv) (p < 0.0001). Quantitative analysis revealed no statistically significant difference between the two groups with respect to signal-to-noise ratio (of pulmonary artery and aorta) and contrast-to-noise ratio. Subjective image quality was significantly better with target mode than with ungated mode (p < 0.0001). CONCLUSION. Target mode prospectively ECG-gated volumetric scanning with iterative reconstruction performed with a 320-MDCT scanner has several benefits in cardiovascular imaging of neonates and infants, including low radiation dose, improved image quality, high diagnostic accuracy, and ability to perform free-breathing studies.Keywords
This publication has 33 references indexed in Scilit:
- Experience with volumetric (320 rows) pediatric CTEuropean Journal of Radiology, 2013
- Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort studyThe Lancet, 2012
- Initial experience on the application of 320-row CT angiography with low-dose prospective ECG-triggered in children with congenital heart diseaseThe International Journal of Cardiovascular Imaging, 2011
- Cognitive and Behavioral Outcomes After Early Exposure to Anesthesia and SurgeryPediatrics, 2011
- Multidetector CT in children: current concepts and dose reduction strategiesPediatric Radiology, 2010
- State-of-the-Art CT Imaging Techniques for Congenital Heart DiseaseKorean Journal of Radiology, 2010
- Early Exposure to Anesthesia and Learning Disabilities in a Population-based Birth CohortAnesthesiology, 2009
- How to use a prospective gated technique for cardiac CTJournal of Cardiovascular Computed Tomography, 2009
- Strategies for Reducing Radiation Dose in CTRadiologic Clinics of North America, 2009
- A Coefficient of Agreement for Nominal ScalesEducational and Psychological Measurement, 1960