Coronary Sinus Lead Extraction
Authors:
- 1 January 2003
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 26 (1p2) , 524-526
Abstract
TYERS, G.F.O., et al.: Coronary Sinus Lead Extraction. Complications are reported more frequently with the implantation of coronary sinus (CS) than other types of leads, and attempts to extract CS leads may also be associated with increased risks. The authors have performed nonthoracotomy lead extraction (LE) since 1981 and maintained a detailed database. By November 2001, 796 leads had been removed from 401 patients. We undertook review of our CS‐LE experience to evaluate prevalence, safety, and efficacy. Of 14 patients referred for CS‐LE, 7 were treated in the last year. In six the lead had been placed in the CS intentionally, and in eight inadvertently. One recent patient treated with biventricular pacing was septic and died before LE was undertaken. In nine men and four women (mean age 66 years) had one each CS lead and a total of 34 LEs (2.6/patient). Four CS leads had been in place forKeywords
This publication has 2 references indexed in Scilit:
- Dissection of the Coronary Sinus Secondary to Pacemaker Lead ManipulationPacing and Clinical Electrophysiology, 2000
- Five‐Years Experience with Intravascular Lead ExtractionPacing and Clinical Electrophysiology, 1994