Anterior vs. posterior position of dispersive patch during radiofrequency catheter ablation: insights fromin silicomodelling
Open Access
- 13 January 2023
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 25 (3) , 1135-1143
Abstract
Aims: To test the hypothesis that the dispersive patch (DP) location does not significantly affect the current distribution around the catheter tip during radiofrequency catheter ablation (RFCA) but may affect lesions size through differences in impedance due to factors far from the catheter tip.Methods: An in silico model of RFCA in the posterior left atrium and anterior right ventricle was created using anatomic measurements from patient thoracic computed tomography scans and tested the effect of anterior vs. posterior DP locations on baseline impedance, myocardial power delivery, radiofrequency current path, and predicted lesion size.Results: For posterior left atrium ablation, the baseline impedance, total current delivered, current distribution, and proportion of power delivered to the myocardium were all similar with both anterior and posterior DP locations, resulting in similar RFCA lesion sizes (< 0.2 mm difference). For anterior right ventricular (RV) ablation, an anterior DP location resulted in slightly higher proportion of power delivered to the myocardium and lower baseline impedance leading to slightly larger RFCA lesions (0.6 mm deeper and 0.8 mm wider).Conclusions: An anterior vs. posterior DP location will not meaningfully affect RFCA for posterior left atrial ablation, and the slightly larger lesions predicted with anterior DP location for anterior RV ablation are of unclear clinical significance.Keywords
Funding Information
- Agencia Nacional de Promoción Científica y Tecnológica
- CONICET
This publication has 21 references indexed in Scilit:
- Effect of Baseline Impedance on Ablation Lesion DimensionsCirculation: Arrhythmia and Electrophysiology, 2018
- Prospective Multicenter Experience With Cooled Radiofrequency Ablation Using High Impedance Irrigant to Target Deep Myocardial Substrate Refractory to Standard AblationJACC: Clinical Electrophysiology, 2018
- Frontal placement of dispersive patch for effective ablation of arrhythmia originating from the anterior right ventricular outflow tractJournal of Interventional Cardiac Electrophysiology, 2017
- Retrograde Coronary Venous Ethanol Infusion for Ablation of Refractory Ventricular TachycardiaCirculation: Arrhythmia and Electrophysiology, 2016
- Bipolar irrigated radiofrequency ablation: A therapeutic option for refractory intramural atrial and ventricular tachycardia circuitsHeart Rhythm, 2012
- Adenosine-Sensitive (Outflow Tract) Ventricular Tachycardia2012
- Subxiphoid Surgical Approach for Epicardial Catheter-Based Mapping and Ablation in Patients With Prior Cardiac Surgery or Difficult Pericardial AccessCirculation, 2004
- Effect of Skin Electrode Location on Radiofrequency Ablation Lesions: An In Vivo and a Three‐Dimensional Finite Element StudyJournal of Cardiovascular Electrophysiology, 1998
- Effects of dispersive electrode position and surface area on electrical parameters and temperature during radiofrequency catheter ablationThe American Journal of Cardiology, 1996
- Anatomy of the Subcutaneous Tissue of the Trunk and Lower ExtremityPlastic and Reconstructive Surgery, 1987