Using Laser-Based Optical Distribution to Achieve First Subpicosecond Multimodule Synchronization in Positron Emission Tomography System
- 4 July 2025
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 74 (00189456) , 1-9
Abstract
As the time performance of the PET detector itself continues to improve, the impact of multi-module synchronization precision on system-level time resolution becomes increasingly evident. Therefore, we propose a laser-based optical distribution synchronization method to achieve higher synchronization precision. Experiments were conducted based on the design, verifying that after the laser carries the reference signal, the additional time uncertainty is only 0.836 ps full width at half maximum (FWHM), and the output signal's slew rate is over six times that of traditional solutions. After optical distribution, the synchronization precision between synchronous signals is 0.855 ps FWHM, with a delay variation range of 0.47 ps. This is an order of magnitude improvement over the precision and stability currently achievable. Moreover, the synchronization precision after 15 km of transmission is 2.34 ps FWHM. The laser-based synchronization method can improve the synchronization precision in the PET system and optimize its integration, and also has the potential to be valuable in ultra-large systems.Keywords
Funding Information
- National Natural Science Foundation of China (12075267)
- Scientific Research Instruments Development Project of Chinese Academy of Sciences (YJKYYQ20200015)
This publication has 58 references indexed in Scilit:
- Optical Properties Modulation: A New Direction for the Fast Detection of Ionizing Radiation in PET2022
- Test and simulation of a Cherenkov picosecond timing counterNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2020
- Timing performance of a Micro-Channel-Plate Photomultiplier TubeNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2020
- Achieving 10 ps coincidence time resolution in TOF‐PET is an impossible dreamMedical Physics, 2020
- Diagnosis of glioma recurrence using multiparametric dynamic 18F-fluoroethyl-tyrosine PET-MRIEuropean Journal of Radiology, 2018
- Precise rise and decay time measurements of inorganic scintillators by means of X-ray and 511 keV excitationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2018
- Introduction2016
- FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0European Journal of Nuclear Medicine and Molecular Imaging, 2014
- Study of TOF PET using Cherenkov lightNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2011
- Functional imaging of neuroendocrine tumors with combined PET/CT using 68Ga‐DOTATATE (DOTA‐DPhe1,Tyr3‐octreotate) and 18F‐FDGCancer, 2008