Monitoring the Bragg peak location of 73 MeV/u carbon ions by means of prompt $γ$-ray measurements
Authors:
- E. TestaETE. Testa
- M. BajardMBM. Bajard
- M. ChevallierMCM. Chevallier
- D. DauvergneDDD. Dauvergne
- F. Le FoulherFFF. Le Foulher
- J. -C. PoizatJPJ. -C. Poizat
- C. RayCRC. Ray
- M. TestaMTM. Testa
- N. FreudNFN. Freud
- J. -M. LétangJLJ. -M. Létang
Preprint
Abstract
By means of a time-of-flight technique, we measured the longitudinal profile of prompt $\gamma$-rays emitted by 73 MeV/u $^{13}$C ions irradiating a PMMA target. This technique allowed us to minimize the shielding against neutrons and scattered $\gamma$-rays, and to correlate prompt gamma emission to the ion path. This correlation, together with a high counting rate, paves the way toward real-time monitoring of the longitudinal dose profile during ion therapy treatments. Moreover, the time correlation between the prompt gamma detection and the transverse position of the incident ions measured by a beam monitor can provide real-time 3D control of the irradiation.Keywords