Abstract
Introduction. An accurate urine analysis is a good indicator of the status of the renal and genitourinary system. However, limited studies have been done on comparing the diagnostic performance of the fully automated analyser and manual urinalysis especially in Ghana. This study evaluated the concordance of results of the fully automated urine analyser (Sysmex UN series) and the manual method urinalysis at the Komfo Anokye Teaching Hospital in Kumasi, Ghana. Methodology. Sixty-seven (67) freshly voided urine samples were analysed by the automated urine analyser Sysmex UN series and by manual examination at Komfo Anokye Teaching Hospital, Ghana. Kappa and Bland-Altman plot analyses were used to evaluate the degree of concordance and correlation of both methods, respectively. Results. Substantial ( , ), slight ( , ), and slight ( , ) agreements were found for urine colour, appearance, and pH, respectively, between the manual and automated methods. A strong and significant correlation ( , ) was found between both methods for specific gravity with a strong positive linear correlation observed for red blood cell count ( , , ), white blood cell count ( , , ), and epithelial cell count (<svg xmlns:xlink="http://www.w3.org/1999/xlink" xmlns="http://www.w3.org/2000/svg"...