Automatic correction of echo‐planar imaging (EPI) ghosting artifacts in real‐time interactive cardiac MRI using sensitivity encoding
Open Access
- 29 November 2007
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 27 (1) , 239-245
Abstract
Purpose To develop a method that automatically corrects ghosting artifacts due to echo-misalignment in interleaved gradient-echo echo-planar imaging (EPI) in arbitrary oblique or double-oblique scan planes. Materials and Methods An automatic ghosting correction technique was developed based on an alternating EPI acquisition and the phased-array ghost elimination (PAGE) reconstruction method. The direction of k-space traversal is alternated at every temporal frame, enabling lower temporal-resolution ghost-free coil sensitivity maps to be dynamically estimated. The proposed method was compared with conventional one-dimensional (1D) phase correction in axial, oblique, and double-oblique scan planes in phantom and cardiac in vivo studies. The proposed method was also used in conjunction with two-fold acceleration. Results The proposed method with nonaccelerated acquisition provided excellent suppression of ghosting artifacts in all scan planes, and was substantially more effective than conventional 1D phase correction in oblique and double-oblique scan planes. The feasibility of real-time reconstruction using the proposed technique was demonstrated in a scan protocol with 3.1-mm spatial and 60-msec temporal resolution. Conclusion The proposed technique with nonaccelerated acquisition provides excellent ghost suppression in arbitrary scan orientations without a calibration scan, and can be useful for real-time interactive imaging, in which scan planes are frequently changed with arbitrary oblique orientations. J. Magn. Reson. Imaging 2007.Keywords
This publication has 14 references indexed in Scilit:
- Phased array ghost eliminationNMR in Biomedicine, 2006
- 32‐element receiver‐coil array for cardiac imagingMagnetic Resonance in Medicine, 2006
- Removal of EPI Nyquist ghost artifacts with two‐dimensional phase correctionMagnetic Resonance in Medicine, 2004
- Real‐time cardiac MRI at 3 teslaMagnetic Resonance in Medicine, 2004
- Multishot EPI-SSFP in the HeartMagnetic Resonance in Medicine, 2002
- Adaptive sensitivity encoding incorporating temporal filtering (TSENSE)†Magnetic Resonance in Medicine, 2001
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Gradient-echo shifting in fast MRI techniques (ERASE imaging) for correction of field inhomogeneity errors and chemical shiftJournal of Magnetic Resonance (1969), 1992
- Echo‐planar imaging with asymmetric gradient modulation and inner‐volume excitationMagnetic Resonance in Medicine, 1990
- Multi-planar image formation using NMR spin echoesJournal of Physics C: Solid State Physics, 1977