A Monte Carlo simulation study on the image resolution in scanning electron microscopy
Authors:
- 29 March 2010
- journal article
- research article
- Published by Wiley in Surface and Interface Analysis
- Vol. 42 (6-7) , 1096-1099
Abstract
The resolution of the scanning electron microscope limited by the beam–sample interaction is studied by Monte Carlo simulation. The simulation model is based on the full Penn algorithm for electron inelastic scattering and the associated secondary electron generation as well as on a constructive solid geometry modeling of a complex sample structure. The line‐scan profiles for gold nanoparticles of different diameters on a carbon substrate are obtained in the simulation. The gap method and contrast‐to‐gradient (C–G) method performed for two separated particles of the same size and different sizes are considered to determine the resolution. Both methods are applied to the simulated images. A comparison on the obtained resolutions performed for the different sample topographies shows that the gap resolution depends excessively on the topography by the shading effect and gap method underestimates the influence of probe size, while C–G method overestimates this influence. Copyright © 2010 John Wiley & Sons, Ltd.Keywords
This publication has 20 references indexed in Scilit:
- Erratum: “Electron inelastic scattering and secondary electron emission calculated without the single-pole approximation” [J. Appl. Phys. 104, 114907 (2008)]Journal of Applied Physics, 2009
- Electron inelastic scattering and secondary electron emission calculated without the single pole approximationJournal of Applied Physics, 2008
- Simulated SEM images for resolution measurementScanning, 2008
- Probe current, probe size, and the practical brightness for probe forming systemsJournal of Vacuum Science & Technology B: Microelectronics and Nanometer Structures Processing, Measurement, and Phenomena, 2008
- Evaluation of both image resolution and contrast-to-noise ratio in scanning electron microscopyJournal of Electron Microscopy, 2007
- Effect of shot noise and secondary emission noise in scanning electron microscope imagesScanning, 2004
- Metrics of resolution and performance for CD-SEMs2000
- Scanning Electron Microscopy1998
- Theoretical study of the ultimate resolution of SEMJournal of Microscopy, 1989
- Quantitative analysis of resolution and stability in nanometer electron beam lithographyJournal of Vacuum Science & Technology B, 1988