Detection of malignant cells in serous body fluids by counting high‐fluorescent cells on the Sysmex XN‐2000 hematology analyzer
- 13 June 2015
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 37 (5) , 715-722
Abstract
Summary: Introduction: The body fluid mode of the Sysmex XN‐2000 hematology analyzer differentiates cells into mononuclear and polymorphonuclear white blood cells (WBC) and high‐fluorescent cells (HFC). The aim of this study was to evaluate the performance of the HFC count for detecting malignant cells in serous body fluids.Methods: Two‐hundred and thirty serous fluids were analyzed on the Sysmex XN body fluid mode. HFC were measured as relative count (HFC/100 WBC) and absolute count (HFC/μL). All samples were microscopically screened on cytospin slides for the presence of malignant cells.Results: Malignant cells were found by microscopic examination in 49 of 230 samples (21.3%). Malignant samples contained significantly higher percentages (10.2 vs. 2.6/100 WBC) and absolute numbers (65 vs. 10/μL) of HFC than nonmalignant samples (P < 0.001). Areas under the ROC curve for relative and absolute HFC count were 0.69 and 0.77, respectively.A cutoff level of ≥17 HFC/μL showed the best performance to predict malignancy, with 88% sensitivity and 61% specificity.Conclusion: As serous body fluids will be more analyzed on automated analyzers in the future, HFC count can be a useful tool to select samples for microscopic review. Microscopic evaluation should be performed if HFC values are above a certain threshold (e.g. ≥17 HFC/μL) or in case of clinical suspicion of malignancy.Keywords
This publication has 12 references indexed in Scilit:
- Performance evaluation of the new hematology analyzer Sysmex XN‐seriesInternational Journal of Laboratory Hematology, 2014
- UF-1000i: validation of the body fluid mode for counting cells in body fluidsClinical Chemistry and Laboratory Medicine (CCLM), 2014
- Automated Evaluation of Synovial and Ascitic Fluids With the Advia 2120 Hematology AnalyzerAmerican Journal of Clinical Pathology, 2013
- Cell counting of body fluids: comparison between three automated haematology analysers and the manual microscope methodInternational Journal of Laboratory Hematology, 2013
- Analysis of serous body fluids using the CELL-DYN Sapphire hematology analyzerClinical Chemistry and Laboratory Medicine (CCLM), 2012
- Performance evaluation of the Sysmex haematology XN modular systemJournal of Clinical Pathology, 2012
- Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluidsClinical Chemistry and Laboratory Medicine (CCLM), 2012
- Automated vs. manual cerebrospinal fluid cell counts: a work and cost analysis comparing the Sysmex XE-5000 and the Fuchs-Rosenthal manual counting chamberInternational Journal of Laboratory Hematology, 2011
- British Thoracic Society Pleural Disease Guidelines - 2010 updateThorax, 2010
- Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluidsClinical Chemistry and Laboratory Medicine (CCLM), 2010