Light output measurements and computational models of microcolumnar CsI scintillators for x‐ray imaging
- 12 January 2015
- journal article
- radiation imaging-physics
- Published by Wiley in Medical Physics
- Vol. 42 (2) , 600-605
Abstract
Purpose: The authors report on measurements of light output and spatial resolution of microcolumnar CsI:Tl scintillator detectors for x‐ray imaging. In addition, the authors discuss the results of simulations aimed at analyzing the results of synchrotron and sealed‐source exposures with respect to the contributions of light transport to the total light output.Methods: The authors measured light output from a 490‐μm CsI:Tl scintillator screen using two setups. First, the authors used a photomultiplier tube (PMT) to measure the response of the scintillator to sealed‐source exposures. Second, the authors performed imaging experiments with a 27‐keV monoenergetic synchrotron beam and a slit to calculate the total signal generated in terms of optical photons per keV. The results of both methods are compared to simulations obtained with hybridmantis, a coupled x‐ray, electron, and optical photon Monte Carlo transport package. The authors report line response (LR) and light output for a range of linear absorption coefficients and describe a model that fits at the same time the light output and the blur measurements. Comparing the experimental results with the simulations, the authors obtained an estimate of the absorption coefficient for the model that provides good agreement with the experimentally measured LR. Finally, the authors report light output simulation results and their dependence on scintillator thickness and reflectivity of the backing surface.Results: The slit images from the synchrotron were analyzed to obtain a total light output of 48 keV−1 while measurements using the fast PMT instrument setup and sealed‐sources reported a light output of 28 keV−1. The authors attribute the difference in light output estimates between the two methods to the difference in time constants between the camera and PMT measurements. Simulation structures were designed to match the light output measured with the camera while providing good agreement with the measured LR resulting in a bulk absorption coefficient of 5 × 10−5 μm−1.Conclusions: The combination of experimental measurements for microcolumnar CsI:Tl scintillators using sealed‐sources and synchrotron exposures with results obtained via simulation suggests that the time course of the emission might play a role in experimental estimates. The procedure yielded an experimentally derived linear absorption coefficient for microcolumnar Cs:Tl of 5 × 10−5 μm−1. To the author's knowledge, this is the first time this parameter has been validated against experimental observations. The measurements also offer insight into the relative role of optical transport on the effective optical yield of the scintillator with microcolumnar structure.Keywords
This publication has 15 references indexed in Scilit:
- Comparison of functional parameters of CsI:Tl crystals and thick filmsRadiation Measurements, 2013
- hybrid$\scriptsize{\mathrm{MANTIS}}$: a CPU–GPU Monte Carlo method for modeling indirect x-ray detectors with columnar scintillatorsPhysics in Medicine & Biology, 2012
- APENELOPE‐based system for the automated Monte Carlo simulation of clinacs and voxelized geometries—application to far‐from‐axis fieldsMedical Physics, 2011
- A fast, angle‐dependent, analytical model of CsI detector response for optimization of 3D x‐ray breast imaging systemsMedical Physics, 2010
- Low-afterglow CsI:Tl microcolumnar films for small animal high-speed microCTNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2009
- Scintillation efficiency, structure and spatial resolution of CsI(Tl) layersNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006
- MANTIS: combined x-ray, electron and optical Monte Carlo simulations of indirect radiation imaging systemsPhysics in Medicine & Biology, 2006
- Lubberts effect in columnar phosphorsMedical Physics, 2004
- Concentration dependence of the light yield and energy resolution of NaI:Tl and CsI:Tl crystals excited by gamma, soft X-rays and alpha particlesNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2002
- Fundamental limits of scintillator performanceNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1993