The Artificial Bowel Sphincter
- 1 December 2011
- journal article
- research article
- Published by Wolters Kluwer Health in Annals of Surgery
- Vol. 254 (6) , 951-956
Abstract
Objective: A report on the long-term results of a consecutive series of patients implanted with the Acticon Neosphincter. Method: Data were reviewed from a prospective database. From May 1996 to Jan 2010, 52 patients (46 women), mean age 51.5 ± 14.8 years, with severe fecal incontinence for a mean of 10.6 ± 10.5 years, were implanted with 85 devices. All patients had failed conservative management, including 13 with unsuccessful prior surgical treatments. Indications for implantation were sphincter destruction (45), pudendal neuropathy (12), congenital malformation (7), and perineal colostomy (4). Preoperative assessment included anal endosonography, anorectal manometry, and electrophysiologic testing. Incontinence (Wexner) and Quality of Life scores were recorded prior to the procedure and at each follow-up visit, together with annual anal physiology assessments. Cumulative risks of device revision and explantation were evaluated using Kaplan-Meier survival curves. Results: Mean follow-up was 64.3 ± 46.5 months (range, 2–169); 26 patients (50%) required revisions after a mean of 57.7 ± 35.0 months, with 73.1% due to a leaking cuff from a microperforation; 14 patients (26.9%) required definitive explantation after a mean of 14.6 ± 7.9 months, with the majority (42.9%) due to infection; and 9 patients were lost to follow-up. In 35 patients (67.3%) with an activated device, there were significant improvements in both median Wexner (P < 0.0001) and Quality-of-life scores (P = 0.0286). There was a significant difference between preoperative resting anal pressures and closed pressures at activation (P < 0.0001) and latest follow-up (P < 0.0001). Conclusion: With careful patient selection, meticulous surgical technique, and dedicated surveillance, favorable long-term results can be achieved with acceptable rates of revision and explantation.Keywords
This publication has 27 references indexed in Scilit:
- Sacral Nerve Stimulation for Fecal IncontinenceAnnals of Surgery, 2010
- Factors Associated with Failure of the Artificial Bowel Sphincter: A Study of Over 50 Cases from Cleveland Clinic FloridaDiseases of the Colon & Rectum, 2009
- Obstetric Anal Sphincter InjuryAnnals of Surgery, 2008
- Systematic review of sacral nerve stimulation for faecal incontinence and constipationBritish Journal of Surgery, 2004
- Artificial Anal Sphincter in Severe Fecal IncontinenceAnnals of Surgery, 2003
- Prevalence of faecal incontinence: obstetric and constipation‐related risk factors; a population‐based studyColorectal Disease, 1999
- Anal Dynamic Graciloplasty in the Treatment of Intractable Fecal IncontinenceNew England Journal of Medicine, 1995
- Etiology and management of fecal incontinenceDiseases of the Colon & Rectum, 1993
- Treatment of Anal Incontinence by an Implantable Prosthetic Anal SphincterAnnals of Surgery, 1992
- IMPLANTATION OF ARTIFICIAL SPHINCTER FOR ANAL INCONTINENCEThe Lancet, 1987