Analytical accuracy of hydrogen measurement using gas chromatography with thermal conductivity detection
Author:
- 22 May 2015
- journal article
- research article
- Published by Wiley in Journal of Separation Science
- Vol. 38 (15) , 2640-2646
Abstract
Gas chromatography employing a thermal conductivity detector with a nitrogen carrier gas and a molecular sieve 5 Å column is commonly used for the analysis of widely varying hydrogen concentrations. Flow variation of the column, caused by carrier gas adsorption, affects the peak shape and impacts the analytical accuracy. The mechanism and factors affecting the adsorption effect are explored, errors caused by the deviation from the linearity of the detector's response are considered, and practical advice is given for improving the analytical accuracy.Keywords
This publication has 10 references indexed in Scilit:
- Thermal Conductivity Detector Analysis of Hydrogen Using Helium Carrier Gas and HayeSep(R) D ColumnsJournal of Chromatographic Science, 1998
- Fundamentals of Gas‐Solid Chromatography1996
- Chromatographic Detectors1996
- Study of peak profiles in nonlinear gas chromatography: Application to the peak distortions observed with overloaded capillary columnsJournal of High Resolution Chromatography, 1985
- Peak distortion in chromatography. Part 2: Kinetically controlled factorsJournal of High Resolution Chromatography, 1982
- Peak distortion in chromatography. Part 1: Concentration‐dependent behaviorJournal of High Resolution Chromatography, 1982
- Gas—modified solid chromatography using organic vapours as carrier gasJournal of Chromatography A, 1974
- Analysis of Hydrogen with Thermal Conductivity DetectorsJournal of Chromatographic Science, 1965
- Adsorption Phenomena and Their Effects on Analytical Accuracy in Gas ChromatographyAnalytical Chemistry, 1961
- Analytical Accuracy in Gas Chromatography Using Thermal Conductivity DetectorsAnalytical Chemistry, 1960