Comparison of the Cytodiff flow cytometric leucocyte differential count system with the Sysmex XE‐2100 and Beckman Coulter UniCel DxH 800
- 18 June 2012
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 34 (6) , 584-593
Abstract
Summary: Introduction: Routine automated haematology analysers categorize leucocytes into five types. The Cytodiff (Beckman Coulter) is a 16‐part leucocyte differential analysis system that uses six markers and five colours. We compared leucocyte differential counts obtained by the Cytodiff with five‐part differential counts obtained by routine automated haematology analysers.Methods: We collected 477 EDTA blood samples from healthy individuals and patients with malignancies, sepsis and multi‐organ failure. Leucocyte differential counts were simultaneously analysed by a Cytodiff multiparametric flow cytometric system and the XE‐2100 (Sysmex) and UniCel DxH 800 (Beckman Coulter) automated haematology analysers. Regression and correlation analyses were performed between the different systems.Results: Our Cytodiff results were well correlated with those produced using the DxH 800 and XE‐2100 analysers except for monocytes and basophils. The correlations were poorer for leukopenic than for nonleukopenic samples. For most samples, Cytodiff obtained a higher correlation with manual counts according to a case analysis; however, in several samples, the Cytodiff generated false decreases in monocyte levels and false increases in basophil levels.Conclusion: The Cytodiff may have an advantage, as it could sensitively detect blasts and immature granulocytes. Additionally, it was less labour‐intensive than manual counting, and therefore, the Cytodiff might be useful for differential counts.Keywords
This publication has 16 references indexed in Scilit:
- Improved Flagging Rates on the Sysmex XE-5000 Compared With the XE-2100 Reduce the Number of Manual Film Reviews and Increase Laboratory ProductivityAmerican Journal of Clinical Pathology, 2011
- Reliable, Accurate Determination of the Leukocyte Differential of Leukopenic Samples by Using Hematoflow MethodAnnals of Laboratory Medicine, 2011
- Granularity Index of the SYSMEX XE-5000 hematology analyzer as a replacement for manual microscopy of toxic granulation neutrophils in patients with inflammatory diseasesClinical Chemistry and Laboratory Medicine (CCLM), 2011
- The new haematology analyzer DxH 800: an evaluation of the analytical performances and leucocyte flags, comparison with the LH 755International Journal of Laboratory Hematology, 2010
- Evaluation of an 8‐color flow cytometric reference method for white blood cell differential enumerationCytometry Part B: Clinical Cytometry, 2010
- Refining the white blood cell differential: The first flow cytometry routine applicationCytometry Part A, 2010
- “6 markers/5 colors” extended white blood cell differential by flow cytometryCytometry Part A, 2007
- Evaluation of the Abbott Cell-Dyn Sapphire Hematology AnalyzerAnnals of Laboratory Medicine, 2007
- Flowcytometric analysis of basophil counts in human blood and inaccuracy of hematology analyzersAllergy, 2005
- Flow Cytometric Validation of Automated Differentials in Pediatric PatientsLaboratory Hematology, 2004