Performance of arthroscopic irrigation systems assessed with automatic blood detection
Open Access
- 9 April 2011
- journal article
- research article
- Published by Wiley in Knee Surgery, Sports Traumatology, Arthroscopy
- Vol. 19 (11) , 1948-1954
Abstract
Purpose: During arthroscopies, bleeding episodes occur as a result of tissue damage. Irrigation systems assist in minimizing these disturbances. The performance of three arthroscopic irrigation systems in clearing bleeding episodes was evaluated objectively.Methods: One surgeon performed 99 shoulder arthroscopies using three irrigation systems: gravity infusion, single roller, and double roller pump. The three irrigation systems groups were matched postoperatively for type of surgery‐acromioplasty, SLAP, release, rotator cuff repair and capsule repair, and operation duration. The recorded arthroscopies were analyzed for the presence of bleeding episodes with a special computer program that automatically detects the tinctures of red‐colored blood. A least 20% of an arthroscopic image had to be covered with blood to qualify as bleeding episode.Results: The median (min–max) presence of bleeding episodes as a percentage of the operation time was 6.6% (0.0–43.6%) for gravity infusion, 3.7% (0.2–46.4%) for the single roller, and 3.3% (0.0–19.3%) for the double roller pump, respectively. The large variation could be attributed to the occurrence of arterial bleeding episodes during some procedures. No significant differences were found between the irrigation systems. For a subgroup including acromioplasties and releases, significant differences were found in favor of both roller pumps (P < 0.05).Conclusions: Overall, the roller pumps did not outperform gravity infusion. However, from the results, high‐risk procedures for bleeding episodes were identified (acromioplasty and release of a frozen shoulder) that can benefit from the use of roller pumps. A clear view is essential to perform an arthroscopic procedure safely and efficiently.Level of evidence: III.Keywords
This publication has 18 references indexed in Scilit:
- Comparison of Operative Times Between Pressure and Flow-Control Pump Versus Pressure-Control Pump for ACL ReconstructionOrthopedics, 2009
- Transferring simulated arthroscopic skills to the operating theatreThe Journal of Bone and Joint Surgery. British volume, 2008
- Haptic Feedback Can Provide an Objective Assessment of Arthroscopic SkillsClinical Orthopaedics and Related Research, 2008
- Comparison of two arthroscopic pump systems based on image qualityKnee Surgery, Sports Traumatology, Arthroscopy, 2008
- Motion Analysis: A Validated Method for Showing Skill Levels in ArthroscopyArthroscopy, 2008
- Drug Infusion Systems2006
- A Controlled Trial of Arthroscopic Surgery for Osteoarthritis of the KneeThe New England journal of medicine, 2002
- Fluid pump systems for arthroscopy: A comparison of pressure control versus pressure and flow controlArthroscopy, 1995
- Three irrigation systems for motorized arthroscopic surgery: A comparative experimental and clinical studyArthroscopy, 1989
- Fluid delivery systems for arthroscopyArthroscopy, 1987