Case Report: Transection of Radial Arterial Catheter Requiring Surgical Intervention
Authors:
Open Access
- 1 January 2018
- journal article
- Published by Scitechnol Biosoft Pvt. Ltd. in Journal of Intensive and Critical Care
- Vol. 04 (01)
Abstract
We report a case in which a radial arterial line was placed prior to induction of general anaesthesia in a 76 year old male with prostate cancer and multiple co-morbidities who presented for thoracoscopic resection of a right lower lobe carcinoid lung tumor. The patient's intra-operative course was complicated by acute blood loss requiring conversion to an open procedure. A subsequent injury to the superior vena cava resulted in the need for immediate re-exploration at the conclusion of surgery. While the arterial line was still indicated in this patient's postoperative course in the ICU, malfunction on post-operative day one resulted in removal of the catheter. The catheter was secured by stay sutures and during the attempt to cut these sutures, the arterial catheter was inadvertently severed and a fragment remained within the patient's radial artery. Surgical intervention was required to remove the retained catheter and the patient recovered without evidence of residual injury or deficit to the artery and extremity. Literature review reveals only one other case report of a retained catheter in a radial artery caused by accidental transection during removal. We review indications for arterial line placement, complications, as well as methods for securing arterial catheters.Keywords
This publication has 16 references indexed in Scilit:
- Surgical and Patient Risk Factors for Severe Arterial Line Complications in AdultsAnesthesiology, 2016
- Iatrogenic catheter sheath shearing during radial artery cannulationKorean Journal of Anesthesiology, 2013
- A retained catheter fragment in radial artery caused by accidental catheter transection during arterial catheter removalJournal of Anesthesia, 2012
- Clinical review: Complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicineCritical Care, 2002
- Indications and complications of arterial catheter use in surgical or medical intensive care units: analysis of 4932 patients.1998
- Another complication of radial artery cannulation.1997
- Locating an arterial foreign body by ultrasonographyBritish Journal of Anaesthesia, 1996
- An iatrogenic arterial foreign bodyBritish Journal of Anaesthesia, 1996
- Effect of Fluids on Life Span of Peripheral Arterial LinesAmerican Journal of Perinatology, 1990
- Complications of radial artery lines in the paediatric patientCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1976