A preliminary study on a novel approach to the assessment of the sample volume length and registration accuracy in PW Doppler quality control
Authors:
- 22 June 2022
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-6
Abstract
Sample volume (SV) in PW Doppler is usually characterized by two main factors, i.e., sample volume size and range gate registration accuracy, which have been included among the recommended Quality Control (QC) Doppler tests commonly proposed in the literature. However, most of the studies focusing on these two factors date back to the last years of the XX century. In this context, the present preliminary study proposes and investigates a novel approach for the automatic analysis of faults in both sample volume length and registration accuracy through the post-processing of PW Doppler spectrograms. In particular, a Velocity Profile Discrepancy Index (VPDI) has been defined and assessed under the hypothesis of laminar flow and parabolic velocity profiles in the examined conduct. According to the definition hereby proposed, VPDI is expected to be as close as possible to 0 for a good agreement between theoretical and measured velocity profile, therefore suggesting that the displayed PW spectrogram is not affected by significant SV error sources. A single ultrasound (US) system of intermediate technology level, equipped with three probe models, has been used to acquire the PW spectrograms obtained for six different SV positions (i.e., six sample volume depths) with respect to the tube diameter of a Doppler flow phantom. Tests have been repeated for all the US probes operating at two Doppler frequencies each. Based on the preliminary promising outcomes, further studies may be needed that include (i) increasing the number of diagnostic systems, (ii) exploring different US system settings, (iii) improving the test protocol to estimate the two error sources separately.Keywords
This publication has 14 references indexed in Scilit:
- A preliminary study on an image analysis based method for lowest detectable signal measurements in Pulsed Wave Doppler ultrasoundsACTA IMEKO, 2021
- Comparative evaluation of three image analysis methods for angular displacement measurement in a MEMS microgripper prototype: a preliminary studyACTA IMEKO, 2021
- A preliminary study on the dynamic characterization of a MEMS microgripper for biomedical applications2021
- A review of Doppler ultrasound quality assurance protocols and test devicesPhysica Medica, 2014
- Simulation and Validation of Arterial Ultrasound Imaging and Blood FlowUltrasound in Medicine & Biology, 2008
- An in vitro system for Doppler ultrasound flow studies in the stenosed carotid artery bifurcationUltrasound in Medicine & Biology, 2002
- Performance tests of Doppler ultrasound equipment with a string phantom.Journal of Ultrasound in Medicine, 1991
- Performance tests of Doppler ultrasound equipment with a tissue and blood-mimicking phantom.Journal of Ultrasound in Medicine, 1988
- Methods to evaluate the sample volume of pulsed doppler systemsUltrasound in Medicine & Biology, 1984
- Technique for studying the sample volume of ultrasonic Doppler devicesMedical & Biological Engineering & Computing, 1973