Is Surface ECG a Useful Surrogate for Subcutaneous ECG?
Authors:
- 1 February 2010
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 33 (2) , 135-145
Abstract
Surface electrocardiograms (ECGs) have been used as surrogates for subcutaneous ECGs to optimize and evaluate subcutaneous devices, but differences between surface and subcutaneous ECGs remain poorly understood. This study evaluated the correspondence between surface and subcutaneous ECGs in Reveal Plus (Medtronic Inc., Minneapolis, MN, USA) patients during various maneuvers. Surface electrodes were placed over the Reveal electrodes of 48 subjects (23 men, age 60 +/- 14.3 years, body mass index 27 +/- 4.9 kg/m(2), implant time 45 +/- 29 weeks). Surface and subcutaneous ECGs were recorded simultaneously for 30 seconds during rest, isometric myopotential noise (pushing palms together), and artifact-inducing maneuvers (repetitive displacement of device, chest thumping on device, arm flaps, handshake, hallwalk). During rest, subcutaneous and surface ECGs were highly correlated (R = 0.96), had similar R-wave amplitude (487 +/- 40 vs 507 +/- 49 microV, NS), and signal-to-noise ratio (SNR) (13.4 +/- 0.8 vs 13.5 +/- 0.7, NS). During myopotential noise, subcutaneous and surface ECGs were highly correlated (R = 0.91) and had similar SNR (10.0 +/- 0.6 vs 9.7 +/- 0.6, NS). During artifact-inducing maneuvers, subcutaneous and surface ECGs were less correlated (R = 0.82 displacement, 0.84 chest thumping, 0.93 arm flaps, 0.90 handshake, 0.92 hallwalk) with subcutaneous SNR significantly higher than surface (11.4 +/- 0.7 vs 9.9 +/- 0.7 displacement, 11.1 +/- 0.6 vs 8.4 +/- 0.6 chest thumping, 11.5 +/- 0.4 vs 10.3 +/- 0.5 arm flaps, 9.5 +/- 0.4 vs 8.4 +/- 0.4 handshake, 12.0 +/- 0.4 vs 10.0 +/- 0.4 hallwalk, P < 0.05). Surface ECGs are adequate surrogates for subcutaneous ECGs in situations free from motion artifacts but not in situations involving movement of the device, surface electrodes, or recording equipment. During artifact-inducing maneuvers, subcutaneous ECGs are of higher quality and less susceptible to artifacts than surface ECGs.Keywords
This publication has 20 references indexed in Scilit:
- Improved Arrhythmia Detection in Implantable Loop RecordersJournal of Cardiovascular Electrophysiology, 2008
- A Detector for a Chronic Implantable Atrial Tachyarrhythmia MonitorIEEE Transactions on Biomedical Engineering, 2008
- Analysing the potential of Reveal® for monitoring cardiac potentialsEP Europace, 2007
- Database testing of a subcutaneous monitor with wireless alarmJournal of Electrocardiology, 2006
- Effect of combined movement and heart rate monitor placement on physical activity estimates during treadmill locomotion and free-livingEuropean Journal of Applied Physiology, 2005
- The feasibility of ST-segment monitoring with a subcutaneous deviceJournal of Electrocardiology, 2004
- Insertable Loop Recorder Use for Detection of Intermittent ArrhythmiasPacing and Clinical Electrophysiology, 2004
- Analysis of electrocardiograms for subcutaneous monitorsJournal of Electrocardiology, 2003
- Implantable Diagnostic Monitoring Devices for Evaluation of Syncope, and Tachy- and Brady-ArrhythmiasJournal of Interventional Cardiac Electrophysiology, 2003
- How Can We Identify the Best Implantation Site for an ECG Event Recorder?Pacing and Clinical Electrophysiology, 2000