A snare in time-adding a safety mechanism to the Amplatzer Septal Occluder
- 16 April 2016
- journal article
- research article
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 87 (5) , 931-932
Abstract
Delivery cable/attachment of the Amplatzer Septal Occluder holds the device at a suboptimal angle relative to the plane of the atrial septum prior to release A small gooseneck snare tightened on the hub can maintain flexible control of the device after release of the delivery cable This technique can improve safety of delivery in difficult cases if you think ahead.Keywords
This publication has 4 references indexed in Scilit:
- Optimal septum alignment of the Figulla® Flex occluder to the atrial septum in patients with secundum atrial septal defectsEuroIntervention, 2016
- Goose‐neck snare‐assisted transcatheter ASD closure: A safety procedure for large and complex ASDsCatheterization and Cardiovascular Interventions, 2016
- A novel snare assistance safeguards against early embolization of devices and facilitates quick retrieval of malpositioned devices in atrial septal defects with deficient marginsAnnals of Pediatric Cardiology, 2015