HistoFlex—a microfluidic device providing uniform flow conditions enabling highly sensitive, reproducible and quantitative in situ hybridizations
- 3 October 2011
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 11 (22) , 3896-3907
Abstract
A microfluidic device (the HistoFlex) designed to perform and monitor molecular biological assays under dynamic flow conditions on microscope slide-substrates, with special emphasis on analyzing histological tissue sections, is presented. Microscope slides were reversibly sealed onto a cast polydimethylsiloxane (PDMS) insert, patterned with distribution channels and reaction chambers. Topology optimization was used to design reaction chambers with uniform flow conditions. The HistoFlex provided uniform hybridization conditions, across the reaction chamber, as determined by hybridization to microscope slides of spotted DNA microarrays when applying probe concentrations generally used in in situ hybridization (ISH) assays. The HistoFlex's novel ability in online monitoring of an in situ hybridization assay was demonstrated using direct fluorescent detection of hybridization to 18S rRNA. Tissue sections were not visually damaged during assaying, which enabled adapting a complete ISH assay for detection of microRNAs (miRNA). The effects of flow based incubations on hybridization, antibody incubation and Tyramide Signal Amplification (TSA) steps were investigated upon adapting the ISH assay for performing in the HistoFlex. The hybridization step was significantly enhanced using flow based incubations due to improved hybridization efficiency. The HistoFlex device enabled a fast miRNA ISH assay (3 hours) which provided higher hybridization signal intensity compared to using conventional techniques (5 h 40 min). We further demonstrate that the improved hybridization efficiency using the HistoFlex permits more complex assays e.g. those comprising sequential hybridization and detection of two miRNAs to be performed with significantly increased sensitivity. The HistoFlex provides a new histological analysis platform that will allow multiple and sequential assays to be performed under their individual optimum assay conditions. Images can subsequently be recorded either in combination or sequentially through the ability of the HistoFlex to monitor assays without disassembly.Keywords
This publication has 39 references indexed in Scilit:
- Fluorescence-Based Codetection with Protein Markers Reveals Distinct Cellular Compartments for Altered MicroRNA Expression in Solid TumorsClinical Cancer Research, 2010
- Characterisation of microRNA expression in post-natal mouse mammary gland developmentBMC Genomics, 2009
- miRNA in situ hybridization in formaldehyde and EDC–fixed tissuesNature Methods, 2009
- A methodology for the combined in situ analyses of the precursor and mature forms of microRNAs and correlation with their putative targetsNature Protocols, 2009
- Use of a multi-thermal washer for DNA microarrays simplifies probe design and gives robust genotyping assaysNucleic Acids Research, 2007
- Detection of microRNAs in frozen tissue sections by fluorescence in situ hybridization using locked nucleic acid probes and tyramide signal amplificationNature Protocols, 2007
- Expression profiling identifies microRNA signature in pancreatic cancerInternational Journal of Cancer, 2006
- Mapping ribosomal RNA transcription activity in the mouse eyeDevelopmental Dynamics, 2006
- In situ detection of miRNAs in animal embryos using LNA-modified oligonucleotide probesNature Methods, 2005
- MicroRNA Gene Expression Deregulation in Human Breast CancerCancer Research, 2005