An International Laboratory Comparison Study of Volumetric and Gravimetric Hydrogen Adsorption Measurements

Authors:

Abstract
In order to determine a material's hydrogen storage potential, capacity measurements must be robust, reproducible, and accurate. Commonly, research reports focus on the gravimetric capacity, and often times the volumetric capacity is not reported. Determining volumetric capacities is not as straight‐forward, especially for amorphous materials. This is the first study to compare measurement reproducibility across laboratories for excess and total volumetric hydrogen sorption capacities based on the packing volume. The use of consistent measurement protocols, common analysis, and figure of merits for reporting data in this study, enable the comparison of the results for two different materials. Importantly, the results show good agreement for excess gravimetric capacities amongst the laboratories. Irreproducibility for excess and total volumetric capacities is attributed to real differences in the measured packing volume of the material.
Funding Information
  • Lawrence Berkeley National Laboratory
  • National Laboratory of the Rockies
  • U.S. Department of Energy (DE-AC36-08GO28308)
  • Sandia National Laboratories
  • U.S. Department of Energy
  • Office of Energy Efficiency and Renewable Energy
  • Fuel Cell Technologies Office (DE-AC36-08-GO28308)
  • National Laboratory of the Rockies (DE-AC04-94AL85000)
  • Sandia National Laboratories (DE-AC02-05CH11231)
  • Argonne National Laboratory

This publication has 28 references indexed in Scilit: