Neutron imaging detector based on the muPIC micro-pixel chamber

Authors:
  • JPJ. D. Parker
    J. D. Parker
  • KHK. Hattori
    K. Hattori
  • HFH. Fujioka
    H. Fujioka
  • MHM. Harada
    M. Harada
  • SIS. Iwaki
    S. Iwaki
  • SKS. Kabuki
    S. Kabuki
  • YKY. Kishimoto
    Y. Kishimoto
  • HKH. Kubo
    H. Kubo
  • SKS. Kurosawa
    S. Kurosawa
  • KMK. Miuchi
    K. Miuchi
  • TNT. Nagae
    T. Nagae
  • HNH. Nishimura
    H. Nishimura

Abstract
We have developed a prototype time-resolved neutron imaging detector employing the micro-pixel chamber (muPIC), a micro-pattern gaseous detector, coupled with a field programmable gate array-based data acquisition system for applications in neutron radiography at high-intensity neutron sources. The prototype system, with an active area of 10cm x 10cm and operated at a gas pressure of 2 atm, measures both the energy deposition (via time-over-threshold) and 3-dimensional track of each neutron-induced event, allowing the reconstruction of the neutron interaction point with improved accuracy. Using a simple position reconstruction algorithm, a spatial resolution of 349 +/- 36 microns was achieved, with further improvement expected. The detailed tracking allows strong rejection of background gamma-rays, resulting in an effective gamma sensitivity of 10^-12 or less, coupled with stable, robust neutron identification. The detector also features a time resolution of 0.6 microseconds.