Validation Study of Fractional Exhaled Nitric Oxide Measurements Using a Handheld Monitoring Device
Authors:
- Michelle GillMGMichelle Gill
- Gavin R. GraffGGGavin R. Graff
- Alan J. AdlerAAAlan J. Adler
- Raed A. DweikRDRaed A. Dweik
- 1 January 2006
- journal article
- research article
- Published by Taylor & Francis in Journal of Asthma
- Vol. 43 (10) , 731-734
Abstract
We tested reproducibility of exhaled nitric oxide (FE(NO)) and inter-operator handling when measured with a handheld device, NIOX MINO. We enrolled 20 volunteers using a priori goals of acceptable reproducibility to be mean within-subject standard deviation less than 3 parts per billion (ppb) for FE(NO) measurements less than 30 ppb, and mean coefficient of variation less than 10% for FE(NO) measurements more than 30 ppb. Seventeen subjects with measurements less than 30 ppb displayed a mean standard deviation of 1.15, and 3 subjects with FE(NO) more than 30 ppb had a mean coefficient of variation of 2.4%. We conclude that NIOX MINO demonstrates excellent reproducibility for all ranges of FE(NO).Keywords
This publication has 32 references indexed in Scilit:
- Exhaled Nitric Oxide Levels during Acute Asthma ExacerbationAcademic Emergency Medicine, 2005
- The Aerocrine exhaled nitric oxide monitoring system NIOX is cleared by the US Food and Drug Administration for monitoring therapy in asthmaJournal of Allergy and Clinical Immunology, 2004
- Effect of β2‐agonist treatment and spirometry on exhaled nitric oxide in healthy children and children with asthmaPediatric Pulmonology, 2002
- Exhaled Nitric Oxide – Is It Really a Good Marker of Airway Inflammation in Bronchial Asthma?Respiration, 2000
- A comparison of exhaled nitric oxide and induced sputum as markers of airway inflammationJournal of Allergy and Clinical Immunology, 2000
- Exhaled nitric oxide levels correlate with measures of disease control in asthmaJournal of Allergy and Clinical Immunology, 2000
- Relationship between exhaled nitric oxide and mucosal eosinophilic inflammation in mild to moderately severe asthmaThorax, 2000
- Comparison of exhaled nitric oxide to spirometry during emergency treatment of asthma exacerbations with glucocorticoids in childrenAnnals of Allergy, Asthma & Immunology, 1999
- Corticosteroids decrease exhaled nitric oxide in children with acute asthmaThe Journal of Pediatrics, 1997
- Expired nitric oxide as a marker for childhood asthmaThe Journal of Pediatrics, 1997