High-resolution multiposition Thomson scattering for the TJ-II stellarator
- 1 January 1999
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (1) , 763-767
Abstract
In this article, the recent installed multiposition Thomson scattering (TS) of the TJ-II stellarator is presented. Light from a pulsed ruby laser is scattered by the free electrons of the plasma and led to a spectrometer for spectral analysis. The Doppler broadened spectrum of the scattered light is analyzed with a spectrometer in the Littrow configuration, detected with a GaAsP image intensifier and recorded with two intensified charged coupled device cameras. Values of the electron temperature in the range of 50 eV up to 4 keV can be measured at 160 spatial elements of 2.25 mm along a chord of 360 mm. First results show that the observational error for is close to the expected error of ∼10% at an electron density of using a laser energy of 8 J.
Keywords
This publication has 6 references indexed in Scilit:
- A high resolution multiposition Thomson scattering system for the Rijnhuizen Tokamak ProjectReview of Scientific Instruments, 1997
- Design and operation of the multipulse Thomson scattering diagnostic on DIII-D (invited)Review of Scientific Instruments, 1992
- TJ-II Project: A Flexible Heliac StellaratorFusion Technology, 1990
- The LIDAR Thomson scattering diagnostic on JET (invited)Review of Scientific Instruments, 1988
- Multichannel Thomson scattering systems with high spatial resolution (invited)Review of Scientific Instruments, 1986
- TFTR edge Thomson scattering systemReview of Scientific Instruments, 1986