Mixing and matching causing taper wear
Authors:
- 1 February 2012
- journal article
- case report
- Published by British Editorial Society of Bone & Joint Surgery in The Journal of Bone and Joint Surgery. British volume
- Vol. 94-B (2) , 281-286
Abstract
We report a case of a male patient presenting with bilateral painful but apparently well-positioned and -fixed large-diameter metal-on-metal hip replacements four years post-operatively. Multiple imaging modes revealed a thick-walled, cystic expansile mass in communication with the hip joint (a pseudotumour). Implant retrieval analysis and tissue culture eliminated high bearing wear or infection as causes for the soft-tissue reaction, but noted marked corrosion of the modular neck taper adaptor and corrosion products in the tissues. Therefore, we believe corrosion products from the taper caused by mismatch of the implant components led to pseudotumour formation requiring revision.Keywords
This publication has 24 references indexed in Scilit:
- Accelerating failure rate of the ASR total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2011
- Histological Features of Pseudotumor-like Tissues From Metal-on-Metal HipsClinical Orthopaedics and Related Research, 2010
- Metal-on-Metal Hip Arthroplasty: Does Early Clinical Outcome Justify the Chance of an Adverse Local Tissue Reaction?Clinical Orthopaedics and Related Research, 2010
- Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2010
- Risk factors for inflammatory pseudotumour formation following hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2009
- Metal–Metal Bearing Surfaces in Hip ArthroplastyOrthopedics, 2009
- Presence of corrosion products and hypersensitivity-associated reactions in periprosthetic tissue after aseptic loosening of total hip replacements with metal bearing surfacesActa Biomaterialia, 2009
- Pseudotumours associated with metal-on-metal hip resurfacingsThe Journal of Bone and Joint Surgery. British volume, 2008
- In vitro comparison of frictional torque and torsional resistance of aged conventional gamma-in-nitrogen sterilized polyethylene versus aged highly crosslinked polyethylene articulating against head sizes larger than 32 mmActa Orthopaedica, 2006
- In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy couplingJournal of Biomedical Materials Research, 1993