Development of robust quantitative methods by near-infrared spectroscopy for rapid pharmaceutical determination of content uniformity in complex tablet matrix
Authors:
- 31 March 2009
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 134 (7) , 1405-1415
Abstract
Robust NIR transmission spectroscopic methods have been developed for determination of content uniformity (CU) of pharmaceutical products with a complex tablet matrix. The tablets of interest, formulated with eight components with active drug load of approximately 30% (w/w), are non-film coated, embossed, and round with thickness values of 3.6 and 5.6 mm, for the 125 and 500 mg dosage strength, respectively. The calibration data set contained seven laboratory scale batches of tablets with concentration range of active pharmaceutical ingredients (API) varying from 85 to 115% relative to label claim (LC) as well as four full scale production batches of tablets that included the natural physical variability of tablets. The reference concentration values were established by a high performance liquid chromatographic method. Partial least-squares (PLS) regression method was used to generate the calibration models. The root mean squared error of calibration for 125 and 500 mg was 1.6 and 1.5% in LC, respectively. The calibration models were validated in terms of measurement accuracy, repeatability, precision, robustness and transferability. Robustness assessment involved challenging the model with tablets incorporating variations in hardness, excipient vendors, excipient content and excipient particle size. The methods exhibited excellent measurement accuracy based on 87 batches (ten tablets for each batch) evaluated. The transferability of the developed NIR methods was demonstrated by comparing the NIR CU results associated with the same set of tablets scanned at the development site with those scanned at the production site. The result indicates that the NIR method can be used as a suitable alternative to the HPLC method for rapid tablet CU release test in pharmaceuticals.Keywords
This publication has 40 references indexed in Scilit:
- Robust Calibration Design in the Pharmaceutical Quantitative Measurements with Near-Infrared (NIR) Spectroscopy: Avoiding the Chemometric PitfallsJournal of Pharmaceutical Sciences, 2009
- Comparison of FT-NIR transmission and UV–vis spectrophotometry to follow the mixing kinetics and to assay low-dose tablets containing riboflavinJournal of Pharmaceutical and Biomedical Analysis, 2006
- Construction of universal quantitative models for determination of roxithromycin and erythromycin ethylsuccinate in tablets from different manufacturers using near infrared reflectance spectroscopyJournal of Pharmaceutical and Biomedical Analysis, 2006
- Content uniformity and tablet hardness testing of intact pharmaceutical tablets by near infrared spectroscopyAnalytica Chimica Acta, 2006
- Validation of a near-infrared transmission spectroscopic procedure: Part B: Application to alternate content uniformity and release assay methods for pharmaceutical solid dosage formsJournal of Pharmaceutical and Biomedical Analysis, 2002
- Development of robust calibration models in near infra-red spectrometric applicationsAnalytica Chimica Acta, 2000
- Analytical control of pharmaceutical production steps by near infrared reflectance spectroscopyAnalytica Chimica Acta, 1999
- Quantitation of the active compound and major excipients in a pharmaceutical formulation by near infrared diffuse reflectance spectroscopy with fibre optical probeAnalytica Chimica Acta, 1996
- Vibrational spectrometry for the assessment of active substance in metoprolol tablets: a comparison between transmission and diffuse reflectance near-infrared spectrometryJournal of Pharmaceutical and Biomedical Analysis, 1996
- Near-Infrared Spectroscopy: Applications in the Analysis of Tablets and Solid Pharmaceutical Dosage FormsApplied Spectroscopy Reviews, 1995