Detection of spontaneous pulse using the acceleration signals acquired from CPR feedback sensor in a porcine model of cardiac arrest
Open Access
- 8 December 2017
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 12 (12) , e0189217
Abstract
Reliable detection of return of spontaneous circulation with minimal interruptions of chest compressions is part of high-quality cardiopulmonary resuscitation (CPR) and routinely done by checking pulsation of carotid arteries. However, manual palpation was time-consuming and unreliable even if performed by expert clinicians. Therefore, automated accurate pulse detection with minimal interruptions of chest compression is highly desirable during cardiac arrest especially in out-of-hospital settings. To investigate whether the acceleration (ACC) signals acquired from accelerometer-based CPR feedback sensor can be used to distinguish perfusing rhythm (PR) from pulseless electrical activity (PEA) in a porcine model of cardiac arrest. Cardiac arrest was induced in 49 male adult pigs. ECG, arterial blood pressure (ABP) and ACC waveforms were simultaneously recorded during CPR. 3-second segments containing compression-free signals during chest compression pauses were extracted and only those segments with organized rhythm were used for analysis. PR was defined as systolic arterial pressure >60 mmHg and pulse pressure >10 mmHg, while PEA was defined as an organized rhythm that does not meet the above criteria for PR. Peak correlation coefficient (CCp) of the cross-correlation function between pre-processed ECG and ACC, was used to discriminate PR and PEA. 63 PR and 153 PEA were identified from the total of 1025 extracted segments. CCp was significantly higher for PR as compared to PEA (0.440±0.176 vs. 0.067±0.042, pr = 0.848, p<0.001). The area under the receiver operating characteristic curve, sensitivity, specificity and accuracy were 0.965, 93.6%, 97.5% and 96.7% for the ACC-based automatic spontaneous pulse detection. In this animal model, the ACC signals acquired from an accelerometer-based CPR feedback sensor can be used to detect the presence of spontaneous pulse with high accuracy.Keywords
Funding Information
- Natural Science Foundation Project of Chongqing (cstc2017jcyjBX0053)
This publication has 37 references indexed in Scilit:
- Feasibility of automated rhythm assessment in chest compression pauses during cardiopulmonary resuscitationResuscitation, 2013
- Transthoracic impedance for the monitoring of quality of manual chest compression during cardiopulmonary resuscitationResuscitation, 2012
- Coronary perfusion pressure during external chest compression in pseudo-EMD, comparison of systolic versus diastolic synchronizationResuscitation, 2012
- Chest Compression Fraction Determines Survival in Patients With Out-of-Hospital Ventricular FibrillationCirculation, 2009
- Reliability of pulse palpation by healthcare personnel to diagnose paediatric cardiac arrestResuscitation, 2009
- A swine model of pseudo-pulseless electrical activity induced by partial asphyxiationResuscitation, 2008
- Interruptions of Chest Compressions During Emergency Medical Systems ResuscitationCirculation, 2005
- Compression depth estimation for CPR quality assessment using DSP on accelerometer signalsIEEE Transactions on Biomedical Engineering, 2002
- Competence of health professionals to check the carotid pulseResuscitation, 1998
- Checking the carotid pulse check: diagnostic accuracy of first responders in patients with and without a pulseResuscitation, 1996