Calibration of a proton beam energy monitora)
Authors:
- M. F. MoyersMMM. F. Moyers
- G. B. CoutrakonGCG. B. Coutrakon
- A. GhebremedhinAGA. Ghebremedhin
- K. ShahnaziKSK. Shahnazi
- P. KossPKP. Koss
- E. SandersESE. Sanders
- 9 May 2007
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (6Part1) , 1952-1966
Abstract
Delivery of therapeutic proton beams requires an absolute energy accuracy of to for patch fields and a relative energy accuracy of to for tailoring the depth dose distribution using the energy stacking technique. Achromatic switchyard tunes, which lead to better stability of the beam incident onto the patient, unfortunately limit the ability of switchyard magnet tesla meters to verify the correct beam energy within the tolerances listed above. A new monitor to measure the proton energy before each pulse is transported through the switchyard has been installed into a proton synchrotron. The purpose of this monitor is to correct and/or inhibit beam delivery when the measured beam energy is outside of the tolerances for treatment. The monitor calculates the beam energy using data from two frequency and eight beam position monitors that measure the revolution frequency of the proton bunches and the effective offset of the orbit from the nominal radius of the synchrotron. The new energy monitor has been calibrated by measuring the range of the beam through water and comparing with published range‐energy tables for various energies. A relationship between depth dose curves and range‐energy tables was first determined using Monte Carlo simulations of particle transport and energy deposition. To reduce the uncertainties associated with typical scanning water phantoms, a new technique was devised in which the beam energy was scanned while fixed thickness water tanks were sandwiched between two fixed parallel plate ionization chambers. Using a multitude of tank sizes, several energies were tested to determine the nominal accelerator orbit radius. After calibration, the energy reported by the control system matched the energy derived by range measurements to better than for all nine energies tested between 40 and with an average difference of . A study of different combinations of revolution frequency and radial offsets to test the envelope of algorithm accuracy demonstrated a relative accuracy of for small energy changes between 126 and . These new measurements may serve as a data set for benchmarking range‐energy relationships.Keywords
This publication has 8 references indexed in Scilit:
- EVA space suit proton and electron threshold energy measurements by XCT and range shiftingRadiation Measurements, 2006
- Test of GEANT3 and GEANT4 nuclear models for 160 MeV protons stopping inMedical Physics, 2003
- LLUPTF: Eleven years and beyondAIP Conference Proceedings, 2002
- Treatment planning for heavy-ion radiotherapy: physical beam model and dose optimizationPhysics in Medicine & Biology, 2000
- Aspects of Fast-Ion DosimetryRadiation Research, 2000
- Electron inelastic-scattering cross sections in liquid waterRadiation Physics and Chemistry, 1998
- Energy loss of 70 MeV protons in elementsNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1992
- Energy loss, range, path length, time-of-flight, straggling, multiple scattering, and nuclear interaction probabilityAtomic Data and Nuclear Data Tables, 1982