High surface area membrane introduction mass spectrometry for analysis of volatile and semi‐volatile organic compounds in air
- 2 August 2001
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 15 (17) , 1520-1524
Abstract
The construction, optimization, and testing of a novel geometry for a semi‐permeable polydimethylsiloxane membrane introduction mass spectrometry (MIMS) system is described. A long poly(dimethylsiloxane) capillary membrane (0.30 mm i.d., 0.64 mm o.d., 60 cm long) was positioned coaxially inside polytetrafluoroethylene tubing (0.80 mm i.d. and 1.60 mm o.d.), allowing for a large surface area membrane of relatively small thickness. Helium, flowing counter‐current on the vacuum side of the membrane, swept the permeate into the mass spectrometer. This optimized geometry for permeation MIMS air analysis allows on‐line detection of volatile and semi‐volatile organic compounds of environmental interest, with limits of detection in the ppt range and a linear response over 3 to 4 orders of magnitude. The system is characterized by short rise and fall times, 10 and 48 s, respectively, for toluene. Copyright © 2001 John Wiley & Sons, Ltd.Keywords
Funding Information
- Center for Sensing Science and Technology, Purdue University
- Office of Naval Research
This publication has 11 references indexed in Scilit:
- Analysis of residual solvents in pharmaceuticals with purge-and-membrane mass spectrometryRapid Communications in Mass Spectrometry, 2000
- On-line monitoring by membrane introduction mass spectrometry of chlorination of organics in water. Mechanistic and kinetic aspects of chloroform formationJournal of Mass Spectrometry, 2000
- Membrane introduction Mass Spectrometry: Trends and applicationsMass Spectrometry Reviews, 2000
- On-line monitoring with feedback control of bioreactors using a high ethanol tolerance yeast by membrane introduction mass spectrometryAnalytica Chimica Acta, 1995
- Letter: Membrane inlet mass spectrometry in the study of the release kinetics of volatile active principles in tisanes of vegetable drugsEuropean Journal of Mass Spectrometry, 1995
- The correlation of membrane permselectivity with Hildebrand solubility parametersJournal of Membrane Science, 1994
- Membrane Introduction Mass SpectrometryAnalytical Chemistry, 1991
- Helium-purged hollow fiber membrane mass spectrometer interface for continuous measurement of organic compounds in waterAnalytical Chemistry, 1991
- Membrane mass spectrometry for the direct trace analysis of volatile organic compounds in air and waterAnalytical Chemistry, 1990
- A study of organic compound pervaporation through silicone rubberJournal of Membrane Science, 1990